Prospective clinical evaluation of two combined spinal-epidural kits.

M J Paech1, S F Evans.   

Abstract

One hundred combined spinal epidural kits (Portex, n = 51, Mallinckrodt, n = 49) were evaluated clinically by twelve anaesthetists with respect to performance of the loss-of-resistance syringe, epidural needle and spinal needle, and success in establishing single-segment combined spinal-epidural anaesthesia for caesarean section. Similar components included a plastic loss-of-resistance device, a Tuohy epidural needle, 26 or 27 gauge pencil-point spinal needle and closed end, three-lateral-eye epidural catheter. The Mallinckrodt kit incorporated a "back-eye" design for exit of the spinal needle from the epidural needle. The epidural and the subarachnoid spaces were satisfactorily identified on 98 occasions. Most anaesthetists preferred the Mallinckrodt kit packaging and the Portex loss-of-resistance syringe, although the latter may have been biased by familiarity with this device. The portex spinal needle was more likely to be felt penetrating the dura (P = 0.02) and aspiration of cerebrospinal fluid was more frequently described as easy (P = 0.01). The most common criticisms of both kits were difficulty controlling the spinal needle position after entry into the subarachnoid space and subjectively, a high degree of resistance to injection through the spinal needle.

Entities:  

Mesh:

Year:  1995        PMID: 8787262     DOI: 10.1177/0310057X9502300513

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  1 in total

1.  Paramedian epidural with midline spinal in the same intervertebral space: An alternative technique for combined spinal and epidural anaesthesia.

Authors:  Deepti Saigal; Rama Wason
Journal:  Indian J Anaesth       Date:  2013-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.