Anthropomorphic phantom measurements for the validation of a treatment planning system.

P Dunscombe1, P McGhee, E Lederer.   

Abstract

A series of eight irradiations of an anthropomorphic phantom has been performed as part of a programme of treatment planning system (TPS) validation. The "treatment' configurations used represent a realistic cross section of those carried out routinely in a radiotherapy centre. The dose distributions within the phantom were determined using 75 LiF TLD chips per irradiation and the measured dose distributions were compared with calculations performed on a TPS for three ranges of dose gradient. The accuracy of the technique per measurement point is estimated to be 3% in dose (low dose gradient) and 3 mm in position (high dose gradient). The measurement and data analysis techniques used permit evaluation of a TPS against recently recommended criteria of acceptability although it is noted that is is not possible to isolate TPS performance from factors depending on, for example, open beam data fitting and treatment set up accuracy.

Mesh:

Year:  1996        PMID: 8778822     DOI: 10.1088/0031-9155/41/3/005

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Independent corroboration of monitor unit calculations performed by a 3D computerized planning system.

Authors:  K W Leszczynski; P B Dunscombe
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

2.  Verification of the accuracy of 3D calculations of breast dose during tangential irradiation: measurements in a breast phantom.

Authors:  S M Kirsner; K L Prado; R C Tailor; J A Bencomo
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.