Thermal stability of premature infants during routine care under radiant warmers.

J H Seguin1, R Vieth.   

Abstract

The body temperatures of infants weighing less than 1500 g under radiant warmers during routine care were documented in the first week of life. Ten infants (median gestational age 28 weeks, median birthweight 913 g) were studied. During 30 nursing interventions (mean 9.2 minutes) mean oesophageal and foot temperature changed 0 degrees C and -0.11 degrees C, respectively. A radiant warmer may limit heat loss during interventions because of easy access and rapid radiant warmer responsiveness.

Entities:  

Mesh:

Year:  1996        PMID: 8777664      PMCID: PMC2528538          DOI: 10.1136/fn.74.2.f137

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747