Literature DB >> 877277

Scattered radiation doses to infants and children during EMI head scans.

D G Bhave, C A Kelsey, J Burstein, B G Brogdon.   

Abstract

Scattered radiation doses to the eyes, thyroid and gonads of infants and children undergoing EMI head scans have been measured with thermoluminescent dosimeters. An average skin dose of 1.3 rad per scan was measured. Results are reported relative to incident skin dose and indicate that it is likely that there is no somatic or genetic hazard due to scattered radiation during an EMI head scan.

Entities:  

Mesh:

Year:  1977        PMID: 877277     DOI: 10.1148/124.2.379

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  The contribution of internal scatter to radiation dose during CT scan of the head.

Authors:  M Chopp; L Ewald; M Hartson
Journal:  Neuroradiology       Date:  1981       Impact factor: 2.804

2.  Irradiation dose to the lens of the eye during CT of the head.

Authors:  E Lund; H Halaburt
Journal:  Neuroradiology       Date:  1982       Impact factor: 2.804

  2 in total