Literature DB >> 8772758

Testing different biphasic waveforms and capacitances: effect on atrial defibrillation threshold and pain perception.

G Tomassoni1, K H Newby, M M Kearney, M J Brandon, H Barold, A Natale.   

Abstract

OBJECTIVES: The goal of this study was to compare the effect of different tilts and capacitances for biphasic shocks on atrial defibrillation efficacy and pain threshold.
BACKGROUND: Although biphasic shocks have been shown to be superior to monophasic shocks, the effect of tilt and capacitance on atrial defibrillation success and pain perception has not been studied in patients.
METHODS: Atrial defibrillation threshold (DFT) testing was performed using a right atrial appendage/coronary sinus lead configuration in 38 patients with a history of paroxysmal atrial fibrillation undergoing an invasive electrophysiologic study. Biphasic waveforms with 40%, 50%, 65%, 80%, 30%/50% and 40%/50% were tested randomly in 22 patients (Group 1). In 16 patients (Group 2), a 65% tilt waveform with 50- and 120-microF capacitance was tested. Before sedation, pain sensation was graded by 15 patients in Group 1 after delivery of a 0.5-J shock and by 10 patients in Group 2 after two 1.5-J shocks with 50- and 120-microF capacitance were delivered.
RESULTS: The DFT energy for the 50% tilt waveform was significantly lower than the 65%, 80% and 30%/50% tilt waveforms. The 40%/50% tilt waveform provided slightly lower energy requirements than the 50% tilt waveform. Nine patients (60%) described the 0.5-J shock as very painful, and four (26.6%) complained of slight pain. The 50-microF capacitor lowered energy requirements compared with the 120-microF capacitor. Six patients (60%) perceived the 1.5-J 50-microF capacitor shock as more painful, whereas three (30%) perceived both shocks as equally painful.
CONCLUSIONS: Biphasic waveforms with 50% tilt in both phases and a smaller tilt in the positive phase than that in the negative phase (40%/50%) provided a decrease in energy requirements at atrial DFT. In addition, stored energy was reduced by biphasic shocks with 50-microF capacitance compared with 120-microF capacitance. Despite the reduction in energy requirements, shocks < 1 J continued to be perceived as painful in the majority of patients.

Entities:  

Mesh:

Year:  1996        PMID: 8772758     DOI: 10.1016/0735-1097(96)00220-3

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  7 in total

Review 1.  New devices and hybrid therapies and new devices for treatment of atrial fibrillation.

Authors:  R B Krol; S Saksena; A Prakash
Journal:  J Interv Card Electrophysiol       Date:  2000-01       Impact factor: 1.900

2.  [Not Available].

Authors:  A Heisel; J Jung; B Schubert; U Michel; H Schieffer; R Fries
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1998-02

3.  Effect of electrode configuration and capacitor size on internal atrial defibrillation threshold using leads currently used for ventricular defibrillation.

Authors:  R Neri; P Palermo; A S Cesario; D Baragli; E Amici; M T Laudadio; A De Rosa; F DeSeta; L Mongeon; G Gambelli
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

Review 4.  Interactions of antiarrhythmic drugs with implantable defibrillator therapy for atrial and ventricular tachyarrhythmias.

Authors:  R B Krol; S Saksena; A Prakash
Journal:  Curr Cardiol Rep       Date:  1999-11       Impact factor: 2.931

5.  Absence of an impact of emotional distress on the perception of intracardiac shock discharges.

Authors:  Karl-Heinz Ladwig; Birgitt Marten-Mittag; Günter Lehmann; Harald Gündel; Heidi Simon; Eckhard Alt
Journal:  Int J Behav Med       Date:  2003

6.  Tetanizing prepulse: A novel strategy to mitigate implantable cardioverter-defibrillator shock-related pain.

Authors:  David W Hunter; Harikrishna Tandri; Henry Halperin; Leslie Tung; Ronald D Berger
Journal:  Heart Rhythm       Date:  2016-01-06       Impact factor: 6.343

Review 7.  Innovations in Clinical Cardiac Electrophysiology: Challenges and Upcoming Solutions in 2018 and Beyond.

Authors:  Vaibhav R Vaidya; Alan Sugure; Samuel J Asirvatham
Journal:  J Innov Card Rhythm Manag       Date:  2017-12-15
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.