Literature DB >> 8762787

Unreamed intramedullary tibial nailing--fatigue of locking bolts.

U W Boenisch1, P G de Boer, S F Journeaux.   

Abstract

Between January 1990 and October 1993 we investigated 72 consecutive patients who had an unreamed tibial nail inserted for fractures of the tibial shaft; 75 per cent (N = 54) were closed fractures and 25 per cent (N = 18) were open fractures. The mean follow up was 15.2 months. Of the patients 91.7 per cent (N = 66) were interlocked, 58 per cent (N = 38) statically and 42 per cent (N = 28) dynamically. Union rate was 98.6 per cent, time to union was 18.5 weeks, 17.3 weeks for closed fractures, 22 weeks for open fractures. Shortening greater than 1 cm occurred in 1.4 per cent (N = 1), varus and valgus deformity in 4.1 per cent each (N = 3). There was no rotational deformity. Of the locking bolts 30 per cent broke 8-10 weeks after partial weight bearing was allowed. Electron microscopy investigations showed that failure of the bolts was due to fatigue. It was not associated with any clinical problems with regard to union times or incidence of malunion. Bolt failure is however a problem if you try to remove the nail.

Entities:  

Mesh:

Year:  1996        PMID: 8762787     DOI: 10.1016/0020-1383(95)00150-6

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  11 in total

1.  Critical analysis of tibial fracture healing following unreamed nailing.

Authors:  Khaled Hamed Salem
Journal:  Int Orthop       Date:  2012-03-17       Impact factor: 3.075

2.  Effect of Locking Plate Fixation vs Intramedullary Nail Fixation on 6-Month Disability Among Adults With Displaced Fracture of the Distal Tibia: The UK FixDT Randomized Clinical Trial.

Authors:  Matthew L Costa; Juul Achten; James Griffin; Stavros Petrou; Ian Pallister; Sarah E Lamb; Nick R Parsons
Journal:  JAMA       Date:  2017-11-14       Impact factor: 56.272

3.  The healing stages of an intramedullary implanted tibia: A stress strain comparative analysis of the calcification process.

Authors:  Vincenzo Filardi
Journal:  J Orthop       Date:  2015-01-31

4.  Fatigue performance of angle-stable tibial nail interlocking screws.

Authors:  Mark Lenz; Boyko Gueorguiev; Robert Geoff Richards; Thomas Mückley; Gunther Olaf Hofmann; Dankward Höntzsch; Markus Windolf
Journal:  Int Orthop       Date:  2012-08-09       Impact factor: 3.075

5.  Comparison of dynamic versus static fixation of intramedullary nailing in tibial diaphyseal fractures: A systematic review and meta-analysis.

Authors:  Andre Jw Loh; James R Onggo; Jason Hockings; Iswadi Damasena
Journal:  J Clin Orthop Trauma       Date:  2022-07-19

6.  Unreamed intramedullary nailing in distal tibial fractures.

Authors:  Khaled Hamed Salem
Journal:  Int Orthop       Date:  2013-07-28       Impact factor: 3.075

7.  Comment on Mohammed et al.: Intramedullary tibial nailing in distal third tibial fractures: distal locking screws and fracture non-union.

Authors:  Devdatta Suhas Neogi; Baldeep Singh; Ashish Jaiman; Chandra Shekhar Yadav; Hira Lal Nag
Journal:  Int Orthop       Date:  2008-08-30       Impact factor: 3.075

8.  Fatigue strength of common tibial intramedullary nail distal locking screws.

Authors:  Lanny V Griffin; Robert M Harris; Joseph J Zubak
Journal:  J Orthop Surg Res       Date:  2009-04-16       Impact factor: 2.359

9.  UK Fixation of Distal Tibia Fractures (UK FixDT): protocol for a randomised controlled trial of 'locking' plate fixation versus intramedullary nail fixation in the treatment of adult patients with a displaced fracture of the distal tibia.

Authors:  Juul Achten; Nicholas R Parsons; Katie R McGuinness; Stavros Petrou; Sarah E Lamb; Matthew L Costa
Journal:  BMJ Open       Date:  2015-09-18       Impact factor: 2.692

10.  The functional results of tibial shaft fractures treated with intramedullary nail compressed by proximal tube.

Authors:  Ahmet Adnan Karaarslan; Nihat Acar; Hakan Aycan; Erhan Sesli
Journal:  Strategies Trauma Limb Reconstr       Date:  2016-02-02
View more