Literature DB >> 8755758

Frameless stereotaxy with real-time tracking of patient head movement and retrospective patient-image registration.

M J Ryan1, R K Erickson, D N Levin, C A Pelizzari, R L Macdonald, G J Dohrmann.   

Abstract

The accuracy of a novel frameless stereotactic system was determined during 10 surgeries performed to resect brain tumors. An array of three charge-coupled device cameras tracked the locations of infrared light-emitting diodes on a hand-held stylus and on a reference frame attached to the patient's skull with a single bone screw. Patient-image registration was achieved retrospectively by digitizing randomly chosen scalp points with the system and fitting them to a scalp surface model derived from magnetic resonance (MR) images. The reference frame enabled continual correction for patient head movements so that registration was maintained even when the patient's head was not immobilized in a surgical clamp. The location of the stylus was displayed in real-time on cross-sectional and three-dimensional MR images of the head; this information was used to predict the locations of small intracranial lesions. The average distance (and standard deviation) between the actual position of the mass and its stereotactically predicted location was 4.8 +/- 3.5 mm. The authors conclude that frameless stereotaxy can be used for accurate localization of intracranial masses without resorting to using fiducial markers during presurgical imaging and without immobilizing the patient's head during surgery.

Entities:  

Mesh:

Year:  1996        PMID: 8755758     DOI: 10.3171/jns.1996.85.2.0287

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  6 in total

Review 1.  Computer-aided navigation in neurosurgery.

Authors:  P Grunert; K Darabi; J Espinosa; R Filippi
Journal:  Neurosurg Rev       Date:  2003-05       Impact factor: 3.042

Review 2.  Neuronavigation and surgery of intracerebral tumours.

Authors:  P W A Willems; J W Berkelbach van der Sprenkel; C A F Tulleken; M A Viergever; M J B Taphoorn
Journal:  J Neurol       Date:  2006-09-20       Impact factor: 4.849

3.  Study on the clinical application of pulsed DC magnetic technology for tracking of intraoperative head motion during frameless stereotaxy.

Authors:  Olaf Suess; Silke Suess; Sven Mularski; Björn Kühn; Thomas Picht; Stefanie Hammersen; Rüdiger Stendel; Mario Brock; Theodoros Kombos
Journal:  Head Face Med       Date:  2006-04-26       Impact factor: 2.151

4.  Localizing ECoG electrodes on the cortical anatomy without post-implantation imaging.

Authors:  Disha Gupta; N Jeremy Hill; Matthew A Adamo; Anthony Ritaccio; Gerwin Schalk
Journal:  Neuroimage Clin       Date:  2014-08-21       Impact factor: 4.881

5.  Vector field analysis for surface registration in computer-assisted ENT surgery.

Authors:  Georgi Diakov; Wolfgang Freysinger
Journal:  Int J Med Robot       Date:  2019-01-07       Impact factor: 2.547

6.  Multimodal navigated skull base tumor resection using image-based vascular and cranial nerve segmentation: A prospective pilot study.

Authors:  Parviz Dolati; Abdulkerim Gokoglu; Daniel Eichberg; Amir Zamani; Alexandra Golby; Ossama Al-Mefty
Journal:  Surg Neurol Int       Date:  2015-11-19
  6 in total