Three comparatively priced intracranial pressure (ICP) microtransducers are now available, each characterized by the manufacturer as having very low zero drift over long periods, an excellent frequency response, and a low measurement error. The three microtransducers, coded Transducer A (Camino OLM ICP monitor; Camino Laboratories, San Diego, CA), Transducer B (Codman Microsensor ICP Transducer; Codman & Shurtlef Inc., Randolph, MA), and Transducer C (ICP Monitoring Catheter Kit OPX-SD [4F]; InnerSpace Medical, Irvine, CA), were examined in a pressure-flow test rig designed for assessment of hydrocephalus shunts. All three microtransducers compiled with the manufacturers' specifications and gave high-quality readings under test conditions. However, some differences were noted; Transducer C had the lowest 24-hour zero drift (drifts in all transducers were < 0.8 mm Hg). The temperature drift was very low in Transducer B and C, but Transducer A had a significantly higher drift (0.27 mm Hg/degrees C). Transducer A had a static error < 0.3 mm Hg, Transducer B < 2 mm Hg, and Transducer C < 8 mm Hg. Frequency detection in Transducers A and B were very good (bandwidth, > 30 Hz), whereas Transducer C had a limited bandwidth of 20 Hz. Transducer B scored the best overall, but all three scored satisfactorily during bench testing.

Entities:  

Mesh:

Year:  1996        PMID: 8747977     DOI: 10.1097/00006123-199601000-00053

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  16 in total

Review 1.  Neuromonitoring in the intensive care unit. I. Intracranial pressure and cerebral blood flow monitoring.

Authors:  Anuj Bhatia; Arun Kumar Gupta
Journal:  Intensive Care Med       Date:  2007-05-24       Impact factor: 17.440

Review 2.  Management of intracranial hypertension.

Authors:  Sunit C Singhi; Lokesh Tiwari
Journal:  Indian J Pediatr       Date:  2009-05-24       Impact factor: 1.967

Review 3.  Model-based indices describing cerebrovascular dynamics.

Authors:  Georgios V Varsos; Magdalena Kasprowicz; Peter Smielewski; Marek Czosnyka
Journal:  Neurocrit Care       Date:  2014-02       Impact factor: 3.210

4.  NICEM consensus on neurological monitoring in acute neurological disease.

Authors:  Peter J D Andrews; Giuseppe Citerio; Luca Longhi; Kees Polderman; Juan Sahuquillo; Peter Vajkoczy
Journal:  Intensive Care Med       Date:  2008-04-09       Impact factor: 17.440

Review 5.  Management of intracranial hypertension.

Authors:  Leonardo Rangel-Castilla; Leonardo Rangel-Castillo; Shankar Gopinath; Claudia S Robertson
Journal:  Neurol Clin       Date:  2008-05       Impact factor: 3.806

6.  Systematic overestimation of intracranial pressure measured using a Camino pressure monitor.

Authors:  M Czosnyka; S Harland; S Piechnik; P Kirkpatrick; Z Czosnyka; D Menon
Journal:  J Neurol Neurosurg Psychiatry       Date:  1996-10       Impact factor: 10.154

7.  Camino intracranial pressure monitor: prospective study of accuracy and complications.

Authors:  R M Martínez-Mañas; D Santamarta; J M de Campos; E Ferrer
Journal:  J Neurol Neurosurg Psychiatry       Date:  2000-07       Impact factor: 10.154

Review 8.  Monitoring and interpretation of intracranial pressure.

Authors:  M Czosnyka; J D Pickard
Journal:  J Neurol Neurosurg Psychiatry       Date:  2004-06       Impact factor: 10.154

9.  Relationship of vascular wall tension and autoregulation following traumatic brain injury.

Authors:  Georgios V Varsos; Karol P Budohoski; Angelos G Kolias; Xiuyun Liu; Peter Smielewski; Vassilis G Varsos; Peter J Hutchinson; John D Pickard; Marek Czosnyka
Journal:  Neurocrit Care       Date:  2014-10       Impact factor: 3.210

10.  Simultaneous monitoring of static and dynamic intracranial pressure parameters from two separate sensors in patients with cerebral bleeds: comparison of findings.

Authors:  Per Kristian Eide; Sverre Holm; Wilhelm Sorteberg
Journal:  Biomed Eng Online       Date:  2012-09-07       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.