Literature DB >> 8740355

Emboli generation by the Medtronic Maxima hard-shell adult venous reservoir in cardiopulmonary bypass circuits: a preliminary report.

S J Mitchell1, T Willcox, C McDougal, D F Gorman.   

Abstract

Increases in right common carotid artery Doppler ultrasound signals typical of emboli were found in cardiopulmonary bypass patients when the Medtronic Maxima hard-shell adult combined venous and cardiotomy reservoir was operated at reservoir blood volumes near the manufacturer;s recommended minimum of 300 ml. The signals were reduced by increasing the reservoir blood volume. Possible microembolus generation in the top- and bottom-entry versions of this reservoir was investigated using an in vitro circuit and a colour flow Doppler monitor that was interfaced with a microprocessor to count the signals. The reservoir blood volume was progressively lowered in 100 ml increments below 1000 ml, and signals were counted over five minutes at each new level. Signal counts downstream of the reservoir increased exponentially after the volume was decreased below 1000 ml in the bottom-entry version, and 700 ml in the top-entry version. Ultrasonic monitoring, both upstream and downstream of the reservoir, showed that the source of these signals was the reservoir itself, and that recirculation of emboli around the circuit accounted for only a small proportion of the measured increase. Changes in circuit blood prime haematocrit within the range 0.11-0.31 did not alter the signal counts. However, counts were profoundly affected by exposure of the circuit to nitrous oxide via a membrane oxygenator; this showed that the emboli were bubbles. Bubble formation may occur where venous blood enters the reservoir as this results in a fountaining effect when the reservoir volume is low. This effect appears to have been serendipitously reduced, but not eliminated, in the design of the top-entry version. It is recommended that perfusionists should not operate these reservoirs at volumes below 1000 and 700 ml in the bottom- and top-entry versions respectively.

Entities:  

Mesh:

Year:  1996        PMID: 8740355     DOI: 10.1177/026765919601100208

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  13 in total

1.  Microbubble generation and transmission of medtronic's affinity hardshell venous reservoir and collapsible venous reservoir bag: an in-vitro comparison.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2011-09

2.  In vitro evaluation of gaseous microemboli handling of cardiopulmonary bypass circuits with and without integrated arterial line filters.

Authors:  Saifei Liu; Richard F Newland; Phillip J Tully; Sigrid C Tuble; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2011-09

Review 3.  What blood pressure is appropriate for cardiopulmonary bypass and how to get it.

Authors:  Alan Merry
Journal:  J Extra Corpor Technol       Date:  2006-03

4.  In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.

Authors:  Timothy A Dickinson; Jeffrey B Riley; Jeffrey C Crowley; Paul M Zabetakis
Journal:  J Extra Corpor Technol       Date:  2006-09

5.  Arterial bubbles from the venous line.

Authors:  Timothy W Willcox; Simon J Mitchell
Journal:  J Extra Corpor Technol       Date:  2006-09

6.  Vacuum assist: angel or demon CON.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2013-06

7.  The ethics of changing practice: do we cross the line?

Authors:  Alan F Merry
Journal:  J Extra Corpor Technol       Date:  2009-12

8.  Perspective on Cerebral Microemboli in Cardiac Surgery: Significant Problem or Much Ado About Nothing?

Authors:  Simon J Mitchell; Alan F Merry
Journal:  J Extra Corpor Technol       Date:  2015-03

9.  Biofeedback: making the science real.

Authors:  Robert A Baker; Richard F Newland; Jayme Bennetts
Journal:  J Extra Corpor Technol       Date:  2009-12

10.  Safer cardiac surgery.

Authors:  Alan F Merry
Journal:  J Extra Corpor Technol       Date:  2009-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.