K E Sharpless1, J B Thomas, L C Sander, S A Wise.
Abstract
A C30 stationary phase was specifically engineered for carotenoid separations, and carotenoid measurements using this column are compared with those obtained using a somewhat more conventional C18 column. Both methods were used to contribute measurements for the certification of carotenoids in Standard Reference Material 968b, Fat-Soluble Vitamins and Cholesterol in Human Serum. Analytes were extracted from the serum into hexane. Measurements on the C18 column were made using a gradient of acetonitrile, methanol, and ethyl acetate, which is described in detail elsewhere. Measurements on the C30 column were made using a gradient of water, methanol, and methyl tert.-butyl ether.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8738021 DOI: 10.1016/0378-4347(95)00494-7
Source DB: PubMed Journal: J Chromatogr B Biomed Appl ISSN: 1572-6495