Finapres tracking of systolic pressure and baroreflex sensitivity improved by waveform filtering.

P Gizdulich1, B P Imholz, A H van den Meiracker, G Parati, K H Wesseling.   

Abstract

OBJECTIVE: Arterial pressure waveforms distort between brachial and finger arteries, causing differences mainly in systolic pressure. Distortion, reportedly, can be removed by applying a waveform filter to the finger pressure.
DESIGN: We analysed the data from two studies that detected discrepancies in systolic tracking between Finapres and brachial pressures. The first set comprised waveforms of seven volunteers during incremental bicycle exercise to exhaustion and the second set comprised waveforms of eight volunteers during increasing phenylephrine infusion.
METHODS: We applied the filter and compared 1 min averaged unfiltered and waveform-filtered finger and brachial pressures.
RESULTS: During exercise, finger systolic pressure overestimated brachial increasingly, from 7(SD 10) mmHg at rest to 27(17) mmHg at maximal exertion. Differences were reduced by waveform filtering from 3 (SD 9)mmHg at rest to 1 (SD 15)mmHg at maximal exertion. During phenylephrine infusion finger systolic pressure overestimated brachial pressure, but the magnitude of the overestimate decreased from 14 (SD 15)mmHg at baseline to -1(SD 16)mmHg at maximal rate. After waveform filtering overestimation was an almost constant 6(SD 11)mmHg. Median baroreflex sensitivities from brachial, unfiltered and waveform-filtered finger pressure were 5.8, 7.5 and 5.3 ms/mmHg and correlation increased after filtering. The results indicate improved systolic pressure tracking after waveform filtering.
CONCLUSIONS: Finger pressure distortion follows a general pattern correctable by waveform filtering. Waveform filtering allows a 'brachial' view to be obtained from Finapres data.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8728303     DOI: 10.1097/00004872-199602000-00014

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  24 in total

1.  Blood pressure measuring devices: recommendations of the European Society of Hypertension.

Authors:  E O'Brien; B Waeber; G Parati; J Staessen; M G Myers
Journal:  BMJ       Date:  2001-03-03

2.  Myogenic responses occur on a beat-to-beat basis in the resting human limb.

Authors:  Seth T Fairfax; Jaume Padilla; Lauro C Vianna; Seth W Holwerda; Michael J Davis; Paul J Fadel
Journal:  Am J Physiol Heart Circ Physiol       Date:  2014-10-31       Impact factor: 4.733

3.  Sex modulates whole-body sudomotor thermosensitivity during exercise.

Authors:  Daniel Gagnon; Glen P Kenny
Journal:  J Physiol       Date:  2011-10-17       Impact factor: 5.182

4.  Heat stress attenuates the increase in arterial blood pressure during isometric handgrip exercise.

Authors:  Konrad Binder; Daniel Gagnon; Aaron G Lynn; Narihiko Kondo; Glen P Kenny
Journal:  Eur J Appl Physiol       Date:  2012-05-29       Impact factor: 3.078

5.  Muscle metaboreceptors modulate postexercise sweating, but not cutaneous blood flow, independent of baroreceptor loading status.

Authors:  Gabrielle Paull; Sheila Dervis; Ryan McGinn; Baies Haqani; Andreas D Flouris; Narihiko Kondo; Glen P Kenny
Journal:  Am J Physiol Regul Integr Comp Physiol       Date:  2015-09-16       Impact factor: 3.619

6.  Adenosine receptor inhibition attenuates the suppression of postexercise cutaneous blood flow.

Authors:  Ryan McGinn; Naoto Fujii; Brendan Swift; Dallon T Lamarche; Glen P Kenny
Journal:  J Physiol       Date:  2014-03-31       Impact factor: 5.182

7.  Brain systems for baroreflex suppression during stress in humans.

Authors:  Peter J Gianaros; Ikechukwu C Onyewuenyi; Lei K Sheu; Israel C Christie; Hugo D Critchley
Journal:  Hum Brain Mapp       Date:  2011-05-12       Impact factor: 5.038

8.  Immediate post-isometric exercise cardiovascular responses are associated with training-induced resting systolic blood pressure reductions.

Authors:  Gavin R Devereux; Jonathan D Wiles; Reuben Howden
Journal:  Eur J Appl Physiol       Date:  2014-10-12       Impact factor: 3.078

9.  Compressed storage of arterial pressure waveforms by selection of significant points.

Authors:  P M de Graaf; J van Goudoever; K H Wesseling
Journal:  Med Biol Eng Comput       Date:  1997-09       Impact factor: 2.602

10.  Short-Term Changes in Cardiovascular Hemodynamics in Response to Bariatric Surgery and Weight Loss Using the Nexfin® Non-invasive Continuous Monitoring Device: a Pilot Study.

Authors:  Sjaak Pouwels; Bianca Lascaris; Simon W Nienhuijs; Arthur R Bouwman; Marc P Buise
Journal:  Obes Surg       Date:  2017-07       Impact factor: 4.129

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.