E F Hunter, R M McKinney, S E Maddison, D D Cruce.
Abstract
The fluorescent treponemal antibody absorption (FTA-ABS) double-staining procedure was reproducible, comparable to the conventional test, and easy to read. We recommend the use of the FTA-ABS double-staining procedure for microscopes with incident illumination, the 100 x/1.30 oil achromatic objective and the 6.3 x ocular to obtain optimal fluorescence, and the KP560 as a barrier filter to exclude rhodamine emission when fluorescein fluorescence is read. With this system, errors related to poor focusing or failure to visualise treponemes on all smears should be eliminated.Entities:
Mesh:
Substances:
Year: 1979 PMID: 87240 PMCID: PMC1045600 DOI: 10.1136/sti.55.2.105