Literature DB >> 8722923

ST segment monitoring with a derived 12-lead electrocardiogram is superior to routine cardiac care unit monitoring.

B J Drew1, M G Adams, M M Pelter, S F Wung.   

Abstract

BACKGROUND: Prior studies have shown that a derived 12-lead electrocardiogram with a simple electrode configuration is comparable with the standard electrocardiogram for arrhythmia analysis.
METHODS: A prospective, comparative, within subjects design was used to compare the value of the derived 12-lead electrocardiogram with that of routine monitoring of leads V1 and II for detection of transient myocardial ischemia in 250 patients treated for unstable angina or myocardial infarction.
RESULTS: During 11,532 hours of derived 12-lead ST segment monitoring, 55 (22%) of 250 patients had 176 episodes of ischemia. Of the 55 patients with ischemia, 75% reported no chest pain and 64% had no ischemic ST changes with routine monitoring leads. All five patients who developed angiographically confirmed abrupt reocclusion after percutaneous transluminal coronary angioplasty had ischemic ST changes with the derived electrocardiogram (sensitivity, 100%), compared with only two patients with routine monitoring (sensitivity, 40%). Serious complications occurred in 17% of angina patients with ischemic events compared to 3% of those without ischemia. Length of stay in the cardiac care unit was twice as long in angina patients who had ischemic events. In patients with acute myocardial infarction, ischemic events were not associated with a more complicated hospital course; however, length of stay in the cardiac care unit was longer in patients with recurrent ischemia.
CONCLUSIONS: The findings show that derived 12-lead ST monitoring is superior to routine monitoring of leads V1 and II for detecting transient myocardial ischemia. ST monitoring of the derived 12-lead electrocardiogram may identify high-risk patients with unstable angina and provide prognostic information that would not be otherwise available from the usual clinical measures.

Entities:  

Mesh:

Year:  1996        PMID: 8722923

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  4 in total

Review 1.  Cardiac Monitoring in the Emergency Department.

Authors:  Jessica K Zègre-Hemsey; J Lee Garvey; Mary G Carey
Journal:  Crit Care Nurs Clin North Am       Date:  2016-07-02       Impact factor: 1.326

2.  Unplanned transfer from the telemetry unit to the intensive care unit in hospitalized patients with suspected acute coronary syndrome.

Authors:  Michele M Pelter; Denise Loranger; Teri M Kozik; Richard Fidler; Xiao Hu; Mary G Carey
Journal:  J Electrocardiol       Date:  2016-08-26       Impact factor: 1.438

3.  Among Unstable Angina and Non-ST-Elevation Myocardial Infarction Patients, Transient Myocardial Ischemia and Early Invasive Treatment Are Predictors of Major In-hospital Complications.

Authors:  Michele M Pelter; Denise L Loranger; Teri M Kozik; Anita Kedia; Richard P Ganchan; Deborah Ganchan; Xiao Hu; Mary G Carey
Journal:  J Cardiovasc Nurs       Date:  2016 Jul-Aug       Impact factor: 2.083

4.  A research method for detecting transient myocardial ischemia in patients with suspected acute coronary syndrome using continuous ST-segment analysis.

Authors:  Michele M Pelter; Teri M Kozik; Denise L Loranger; Mary G Carey
Journal:  J Vis Exp       Date:  2012-12-28       Impact factor: 1.355

  4 in total