Literature DB >> 8699952

Phase-encode reordering to minimize errors caused by motion.

C K Macgowan1, M L Wood.   

Abstract

A new method for suppressing the effects of motion in MR images by reordering the acquisition of k space has been developed. Existing reordering methods suffer from image blurring. The method presented here applies specifically to translation along the phase-encoding direction, in which case it reduces both ghosting and blurring. The method is conceptually similar to a linear phase shift of k space, except that it has the advantage of not corrupting stationary structures inadvertently. The method is intended for anatomic sites in which substantial translational motion occurs along the phase-encoding direction, such as the cranial-caudal motion of the liver and kidneys. The reordering method is motivated from an analysis of factors affecting the severity of motion artifacts. The theory behind the reordering method is described and validated experimentally by imaging a moving phantom.

Mesh:

Year:  1996        PMID: 8699952     DOI: 10.1002/mrm.1910350318

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  2 in total

1.  MR CAT scan: a modular approach for hybrid imaging.

Authors:  C Hillenbrand; D Hahn; A Haase; P M Jakob
Journal:  MAGMA       Date:  2000-07       Impact factor: 2.310

2.  Comparison of parallel MRI reconstruction methods for accelerated 3D fast spin-echo imaging.

Authors:  Zhikui Xiao; W Scott Hoge; R V Mulkern; Lei Zhao; Guangshu Hu; Walid E Kyriakos
Journal:  Magn Reson Med       Date:  2008-09       Impact factor: 4.668

  2 in total