| Literature DB >> 8697576 |
T Lövgren1, L Meriö, K Mitrunen, M L Mäkinen, M Mäkelä, K Blomberg, T Palenius, K Pettersson.
Abstract
The availability of an intrinsically fluorescent, inert, and stable Eu chelate label made it feasible to design one-step all-in-one immunoassays with time-resolved fluorometry for detection. Both competitive and noncompetitive immunoassays are performed in microtitration wells containing all assay-specific components in a stable dry form. Only the sample and one assay buffer common for all analytes need to be added. Model assays for human chorionic gonadotropin (hCG), alpha-fetoprotein (AFP), and progesterone all reached equilibrium in 15 min or less without compromising the performance characteristics of the measurements, all of which perform at least equivalent to state-of-the-art assays. The detection limits for hCG, AFP, and progesterone were 0.3 IU/L, 0.1 microgram/L, and 0.5 nmol/L, respectively. The assay ranges for hCG and AFP were linear to 5000 IU/L and 1200 micrograms/L, respectively. The immunoassay format can be readily implemented in a fully automated random-access immunoassay system with optimal performance characteristics and no handling of analyte-specific assay components.Entities:
Mesh:
Substances:
Year: 1996 PMID: 8697576
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327