| Literature DB >> 8682534 |
R L Seaman1, J Chen.
Abstract
A sensing platform with two piezoelectric transducers was designed and fabricated to measure acoustic startle responses from forelimbs and hindlimbs in the rat. Testing with a vibrator showed that separate forces were measured from 5 to 25 Hz with mean sensitivities of 2.395 and 2.022 V/N and mean linearity errors of 3.23 and 2.98% FS for the forelimb and hindlimb sensors, respectively. Forelimb and hindlimb response waveforms of male Sprague-Dawley rats had shapes similar to the commonly recorded wholebody response but were smaller in amplitude.Entities:
Mesh:
Year: 1996 PMID: 8682534 DOI: 10.1109/10.481992
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538