A comparison of two transcutaneous monitors for the measurement of arterial PO2 and PCO2 in neonates.

B Carter1, M Hochmann, A Osborne, A Nisbet, N Campbell.   

Abstract

We examined the ability of two transcutaneous devices (Fastrac, Sensormedics Corporation, Yorba Linda, California, U.S.A. and Hewlett Packard M1018A, Hewlett Packard Component Monitoring System, Hewlett Packard, North Hollywood, U.S.A.) to measure arterial PCO2 and PO2 in neonates. Thirty-seven neonates had transcutaneous oxygen measured with the Hewlett Packard (HPO2 group), 38 neonates had transcutaneous carbon dioxide measured with the Hewlett Packard (HPCO2 group) and the Fastrac was used on 27 neonates (FTCO2 group). Both devices were operated with electrode temperatures of 43.5 degrees C although an additional ten subjects were studied using the Fastrac with an electrode temperature of 43.0 degrees C. The mean differences (transcutaneous--arterial) and upper and lower limits of agreement were calculated for each group. For the HPO2 group they were 3.78 mmHg (-12.23 to 19.80 mmHg), for the HPCO2 group they were 0.40 mmHg (-4.50 to 5.30 mmHg) and for the FTCO2 they were -0.96 mmHg (-7.85 to 5.92 mmHg). For the Fastrac group at an electrode temperature of 43.0 degrees C the mean difference and limits of agreement were -1.00 mmHg and -4.58 mmHg to 2.58 mmHg. The average sensitivity and specificity for both machines for the detection of hypocarbia were 82% and 92% respectively while for hypercarbia they were 90% and 94% respectively. For hypoxaemia, the sensitivity and specificity were 40% and 94% while for hyperoxaemia the sensitivity and specificity were 83% and 97%. We conclude that both machines provide a useful supplement to arterial PCO2 measurements and the Fastrac performs better at 43.0 degrees C. The measurement of PO2 is less accurate but is still of clinical use.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8669606     DOI: 10.1177/0310057X9502300610

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  4 in total

Review 1.  Noninvasive monitoring of peripheral perfusion.

Authors:  Alexandre Lima; Jan Bakker
Journal:  Intensive Care Med       Date:  2005-09-17       Impact factor: 17.440

2.  Clinical validation of a digital transcutaneous PCO2/SpO2 ear sensor in adult patients after cardiac surgery.

Authors:  Werner Baulig; Philipp Schütt; Hans R Roth; Josef Hayoz; Edith R Schmid
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

3.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

4.  Monitoring Oxygenation and Gas Exchange in Neonatal Intensive Care Units: Current Practice in the Netherlands.

Authors:  Ratna N G B Tan; Estelle E M Mulder; Enrico Lopriore; Arjan B Te Pas
Journal:  Front Pediatr       Date:  2015-11-03       Impact factor: 3.418

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.