Imaging characteristics of x-ray capillary optics in digital mammography.

D G Kruger1, C C Abreu, E G Hendee, A Kocharian, W W Peppler, C A Mistretta, C A MacDonald.   

Abstract

Computed radiography (CR) has shown promise in digital mammographic screening due to its good low spatial frequency MTF and its relatively wide exposure latitude. The CR image format has not gained acceptance clinically because of reduced high spatial frequency resolution as compared to film-screen images. X-ray capillary optics, aligned between the breast and CR phosphor imaging plate, will capture primary x-ray photons almost exclusively. Due to the very small angle of acceptance, scattered photons angled more than about 1.6 x 10(-3) radians from primary trajectory will not be accepted at the capillary optic entrance. The virtual elimination of detected scatter means almost 100% of the possible primary contrast should be visible in the image. In addition, the image can be magnified without focal spot blurring. Effective resolution of CR images can be increased by a factor equal to that magnification. Clinical implementation of future capillary optics are expected to be either in the form of a large, stationary, post-patient optic that accepts primary from the entire breast or a fan-shaped optic that is scanned across the breast. Measurements of a test capillary optic showed a reduction of scatter fraction to 0.018. Images of a lucite contrast detail phantom revealed a corresponding increase in image contrast when compared to anti-scatter grid and no grid methods. Spectral transmission measurements using a high-purity germanium detector showed good primary transmission (45%-50%) in the mammographic energy range. The MTF measurements of both stationary and scanned capillary optics showed improvement at the 5% MTF level to 8.4 mm-1 for scanned optics and 9.2 mm-1 for stationary optics representing a 68% and 84% respective increase over the CR MTF without magnification or capillary optics.

Entities:  

Mesh:

Year:  1996        PMID: 8668099     DOI: 10.1118/1.597703

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Usefulness of a small-field digital mammographic imaging system using parabolic polycapillary optics as a diagnostic imaging tool: a preliminary study.

Authors:  Kwon Su Chon; Jeong Gon Park; Hyun Hwa Son; Sung Hoon Kang; Seong Hoon Park; Hye-Won Kim; Hun Soo Kim; Kwon-Ha Yoon
Journal:  Korean J Radiol       Date:  2009 Nov-Dec       Impact factor: 3.500

2.  Controlling X-ray beam trajectory with a flexible hollow glass fibre.

Authors:  Yoshihito Tanaka; Takashi Nakatani; Rena Onitsuka; Kei Sawada; Isao Takahashi
Journal:  J Synchrotron Radiat       Date:  2013-12-10       Impact factor: 2.616

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.