Literature DB >> 8666629

Fracture of the metal tibial tray after Kinematic total knee replacement. A common cause of early aseptic failure.

P J Abernethy1, C M Robinson, R M Fowler.   

Abstract

We reviewed 1567 elective knee replacements performed between 1980 and 1990, using either the Total Condylar prosthesis with an all-plastic tibial component, or the Kinematic prosthesis which has a metal tibial tray. The ten-year probability of survival was 92.1% for the Total Condylar design and 87.9% for the Kinematic. The difference was mainly due to 16 revisions required in the Kinematic series for fracture of the metal base-plate. This was the most common cause of aseptic failure in this group. These fractures were strongly associated with a preoperative varus deformity (hazard ratio (HR) 8.8) and there was a slightly increased risk in males (HR 1.9) and in osteoarthritic knees (HR 1.8). In the nine fractures which occurred within four years of primary implantation (group 1), failure to correct adequately a preoperative varus deformity and the use of a bone graft to correct such a deformity were both strongly associated with fracture (HR 13.9 and 15.8, respectively). In eight fractures which occurred more than five years after primary replacement (group 2) we could detect no significant risk factors. Early complications occurred in two patients after the 16 revision procedures for tray fracture. One had a deep infection and the other refracture of the tray.

Entities:  

Mesh:

Year:  1996        PMID: 8666629

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  12 in total

Review 1.  Causes of failure and etiology of painful primary total knee arthroplasty.

Authors:  Romain Seil; Dietrich Pape
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-08-11       Impact factor: 4.342

Review 2.  The incidence of implant fractures after knee arthroplasty.

Authors:  Magdalena M Gilg; Christian W Zeller; Lukas Leitner; Andreas Leithner; Gerold Labek; Patrick Sadoghi
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-05-06       Impact factor: 4.342

3.  Relationship between Tibial Bone Defect and Extent of Medial Release in Total Knee Arthroplasty.

Authors:  Shin Woo Nam; Ji Hoon Kwak; Nam Ki Kim; Il Whan Wang; Beom Koo Lee
Journal:  Knee Surg Relat Res       Date:  2012-09-03

4.  Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.

Authors:  Alan D Lam; Gavan P Duffy
Journal:  Arthroplast Today       Date:  2022-10-12

5.  Bacterial contamination of the wound during primary total hip and knee replacement. Median 13 years of follow-up of 90 replacements.

Authors:  Eythor Örn Jonsson; Hera Johannesdottir; Otto Robertsson; Brynjolfur Mogensen
Journal:  Acta Orthop       Date:  2014-04       Impact factor: 3.717

6.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

7.  Ceramic femoral components in total knee arthroplasty - two year follow-up results of an international prospective multi-centre study.

Authors:  Philipp Bergschmidt; Rainer Bader; Dirk Ganzer; Christian Hauzeur; Christoph Lohmann; Wolfgang Rüther; Domenico Tigani; Nicola Rani; Fernando Lopez Prats; Claudio Zorzi; Vincenzo Madonna; Stefano Rigotti; Francesco Benazzo; Stefano Marco Paolo Rossi; Guenther Kundt; Hans Rudolf Bloch; Wolfram Mittelmeier
Journal:  Open Orthop J       Date:  2012-04-20

8.  Nontraumatic fracture of the femoral condylar prosthesis in a total knee arthroplasty leading to mechanical failure.

Authors:  Girish N Swamy; Conal Quah; Elmunzar Bagouri; Nitin P Badhe
Journal:  Case Rep Orthop       Date:  2014-01-23

9.  FRACTURE OF THE TIBIAL COMPONENT IN TOTAL KNEE ARTHROPLASTY: REPORT ON TWO CASES.

Authors:  Idemar Monteiro da Palma; Rodrigo Pires E Albuquerque; João Maurício Barretto
Journal:  Rev Bras Ortop       Date:  2015-12-08

10.  TIBIAL PERIPROSTHETIC FRACTURE COMBINED WITH TIBIAL STEM STRESS FRACTURE FROM TOTAL KNEE ARTHROPLASTY.

Authors:  Fernando Fonseca; Edgar Rebelo; Antonio Completo
Journal:  Rev Bras Ortop       Date:  2015-11-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.