Comparison of absorbed radiation doses in barium and air enema reduction of intussusception: a phantom study.

J Persliden1, P Schuwert, W Mortensson.   

Abstract

OBJECTIVE: We assessed the relative radiation load in patients undergoing hydrostatic and pneumatic reduction of childhood intussusception.
MATERIALS AND METHODS: In a phantom study we simulated two situations occurring during reduction of intussusception. The absorbed radiation dose was measured at several positions in the phantom using either barium sulphate (BaSO4) or air in the simulated reduction, combined with either automatic exposure control (AEC) or constant exposure rate (CER) at fluoroscopy. From these values the mean absorbed dose was calculated for different depth compartments within the phantom.
RESULTS: In the barium study the mean absorbed dose averaged over the total irradiated volume was 14-23 % lower when CER was used instead of AEC; in the air study the dose was 35-43 % lower when AEC was used instead of CER. The combination of air and AEC provided the lowest mean absorbed dose in the tissue. The barium enema created a low-radiation zone, which might be utilized for protecting radiation sensitive tissue.
CONCLUSION: The use of BaSO4 or air in reduction of intussusception requires the proper combination with CER and AEC, respectively, to minimize the radiation load to the patient; the lowest radiation load is obtained by using air and AEC.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8657461     DOI: 10.1007/bf01395708

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  4 in total

1.  Estimation of doses to patients from "complex" conventional X-ray examinations.

Authors:  A Calzado; E Vañó; P Morán; C Castellote; S Ruiz; L González
Journal:  Br J Radiol       Date:  1991-06       Impact factor: 3.039

2.  Pneumatic reduction: advantages, risks and indications.

Authors:  D A Stringer; S H Ein
Journal:  Pediatr Radiol       Date:  1990

3.  Dosimetry at interfaces. Theoretical analysis and measurements by means of thermoluminescent LiF at plane interfaces between a low Z-material and Al, Cu, Sn or Pb irradiated with 100 to 200 kV roentgen radiation.

Authors:  G A Carlsson
Journal:  Acta Radiol Suppl       Date:  1973

4.  Radiation exposure to children in diagnosing and at hydrostatic reduction of intussusception.

Authors:  A Karlsson; P Schuwert; W Mortensson
Journal:  Acta Radiol       Date:  1994-05       Impact factor: 1.990

  4 in total
  3 in total

1.  Intussusception reduction: Effect of air vs. liquid enema on radiation dose.

Authors:  Summer L Kaplan; Dennise Magill; Marc A Felice; J Christopher Edgar; Sudha A Anupindi; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2017-06-03

2.  Female gonadal shielding with automatic exposure control increases radiation risks.

Authors:  Summer L Kaplan; Dennise Magill; Marc A Felice; Rui Xiao; Sayed Ali; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2017-10-18

3.  Radiation dose in pneumatic reduction of ileo-colic intussusceptions--results from a single-institution study.

Authors:  Jennifer L Cullmann; Johannes T Heverhagen; Stefan Puig
Journal:  Pediatr Radiol       Date:  2014-11-23
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.