Literature DB >> 8653945

Ceramic femoral head fractures in total hip arthroplasty.

E W Fritsch1, M Gleitz.   

Abstract

A failure analysis was performed of 4341 alumina ceramic heads articulating with 2693 alumina ceramic and 1464 polymer sockets implanted over 20 years (1974 to 1994). From 1974 to 1982, a mushroom shaped head with ceramic neck was used in 1069 cases, and from 1982 to 1994 a ball type head was used in 3272 cases. In the ceramic/ceramic cases, the average followup was 11 years, and in the polymer pairing cases, the average followup was 6 years. In ceramic self pairing with the mushroom shaped head, the fracture rate was 0.4% (5 of 1069). With the ball type head, the fracture rate was 0.06% (1 of 1763). In articulation with polymer sockets, only 1 head fracture occurred (0.07%). In the group of cases with ceramic/ceramic pairing, the reason for fracture was direct trauma in 4 cases, recurrent neck impingement in 2 cases, and fatigue failure in 1 case. The only case with ceramic head fracture in polymer pairing also was caused by direct trauma. Fractures of the alumina ceramic heads cannot be avoided, but the use of ball type neckless heads brought the fracture rate close to 0. Under the aspect of material safety, it seems to be possible to use the great advantage of the superior low wear of the alumina/alumina couple with negligible fracture risk.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8653945     DOI: 10.1097/00003086-199607000-00021

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  20 in total

1.  The case of the hip grinders: fracture and failure of ceramic femoral head prostheses.

Authors:  V Sodiwala; V Bg Rao; J Pillay
Journal:  Emerg Med J       Date:  2006-08       Impact factor: 2.740

Review 2.  Manufacturers affect clinical results of THA with zirconia heads: a systematic review.

Authors:  Hiroyuki Yoshitomi; Satoru Shikata; Hiromu Ito; Takeo Nakayama; Takashi Nakamura
Journal:  Clin Orthop Relat Res       Date:  2009-01-31       Impact factor: 4.176

3.  Ideal femoral head size in total hip arthroplasty balances stability and volumetric wear.

Authors:  Michael B Cross; Denis Nam; David J Mayman
Journal:  HSS J       Date:  2012-09-13

4.  11β-hydroxysteroid dehydrogenases as targets in the treatment of steroid-associated femoral head necrosis using antler extract.

Authors:  Ribusurong Pu; Hao Peng
Journal:  Exp Ther Med       Date:  2017-11-07       Impact factor: 2.447

Review 5.  The incidence of implant fractures after total hip arthroplasty.

Authors:  Patrick Sadoghi; Wolfram Pawelka; Michael C Liebensteiner; Alexandra Williams; Andreas Leithner; Gerold Labek
Journal:  Int Orthop       Date:  2014-01       Impact factor: 3.075

6.  Ceramic liner fracture and impingement in total hip arthroplasty.

Authors:  Amy Steinhoff; Victor Hakim; Richard H Walker; Clifford W Colwell; Steven N Copp
Journal:  HSS J       Date:  2014-10-21

7.  Alumina-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty.

Authors:  Yup Lee Jung; Shin-Yoon Kim
Journal:  Open Orthop J       Date:  2010-02-11

8.  Charnley-Kerboull total hip arthroplasty combining zirconia on polyethylene. A minimum eight-year follow-up prospective study.

Authors:  Jean Langlois; Samer El Hage; Faouzi Madi; Jean-Pierre Courpied; Marcel Kerboull; Moussa Hamadouche
Journal:  Int Orthop       Date:  2013-01-31       Impact factor: 3.075

Review 9.  [The problem of fractures of ceramic heads. What should be done?].

Authors:  H R Dürr
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

10.  Porous metals and alternate bearing surfaces in shoulder arthroplasty.

Authors:  Shannon R Carpenter; Ivan Urits; Anand M Murthi
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.