ELISA for thyroglobulin in serum: recovery studies to evaluate autoantibody interference and reliability of thyroglobulin values.

M Erali1, R B Bigelow, A W Meikle.   

Abstract

An ELISA for measuring thyroglobulin (Tg) in serum was developed with polyclonal antibodies to Tg on the solid phase and two monoclonal antibodies to Tg as the second antibodies. The assay has a detection limit of 1 microgram/L, a within-run imprecision (CV) of <7%, and a between-run CV of < 10%. Parallelism of the assay was shown in dilution studies, in which the percent observed/expected values for n = 5 autoantibody-containing samples gave a mean of 99% (SD 13.1 %); for n = 5 samples with undetectable autoantibody concentrations, the mean was 103% (SD 11.8%). The correlation of the ELISA with an RIA for Tg in 46 normal samples was ELISA = 1.11(RIA) + 0.52, S y/x = 2.23, SD intercept = 0.54, SD slope = 0.03, range = 0 to 53 micrograms/L, r = 0.980. Comparison of the ELISA with a reference laboratory RIA for 29 clinical samples gave a correlation of: ELISA = 1.53(RIA) - 0.48, S y/x = 9.00, SD intercept = 2.19, SD slope = 0.10, range = 0 to 98 micrograms/L, r = 0.950. To provide additional information concerning the reliability of the Tg measurement in samples containing autoantibodies to Tg, we developed a procedure for determining the percent recovery. A percent recovery greater than or equal to 80% indicates minimal interference by autoantibodies in this assay. The assay is straightforward to perform, results can be posted within 8 h, and the routinely good recovery of Tg in the presence of Tg autoantibodies indicates minimal autoantibody interference in this assay.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8653905

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

Review 1.  Harmonization of blood-based indicators of iron status: making the hard work matter.

Authors:  Andrew N Hoofnagle
Journal:  Am J Clin Nutr       Date:  2017-10-25       Impact factor: 7.045

2.  Comparison of immunoradiometric assays for determination of thyroglobulin: a validation study.

Authors:  L A Tortajada-Genaro; M P Cózar; J L Moreno Frigols; C Romero de Avila
Journal:  J Clin Lab Anal       Date:  2007       Impact factor: 2.352

3.  Experiences from occupational exposure limits set on aerosols containing allergenic proteins.

Authors:  Gunnar D Nielsen; Søren T Larsen; Jitka S Hansen; Lars K Poulsen
Journal:  Ann Occup Hyg       Date:  2012-07-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.