Collimator-related radiation dose for different cobalt machines and linear accelerators.

P H van der Giessen1.   

Abstract

PURPOSE: In previous publications (12, 13) measurements are described of the dose outside the primary beam (the peripheral dose (PD)) for 60Co gamma radiation to 25 MV photons. Comparison with data published by other investigators for different treatment machines, showed good agreement. This can only be explained when the contribution to the PD of radiation leaking through and scattering from the collimator does not differ considerably between treatment machines from different manufactures, and it is the purpose of this article to investigate whether this assumption is valid. METHODS AND MATERIALS: A request was sent out to all radiotherapy departments in The Netherlands and one in Belgium to measure the dose outside the primary beam for as many machines as possible. The following geometry was given: field sizes of 10 x 10 cm2 and 20 x 20 cm2 at distances of 30 and 50 cm, for collimator angles 0 degrees and 90 degrees at the standard source surface distance. This, therefore, resulted in a dataset of eight measurements per photon energy.
RESULTS: Data were collected for four cobalt machines and 37 linear accelerators, from seven different manufacturers. All together 56 datasets were collected for 12 different photon energies. Although the variation of the leakage radiation dose is small, there can be differences of about 50% in the collimator scatter dose between collimator angles of 0 degrees and 90 degrees, depending on the collimator design or on the design of the flattening filter. For dual energy machines with a large gap between the low and the high energy, the values for the high energy are higher by about 40%. Old cobalt machines show higher leakage radiation dose than modern ones.
CONCLUSION: Although there is no large variation in leakage radiation dose between different makes of accelerators, some show higher collimator scatter dose than others. The magnitude of the leakage radiation dose is well within regulatory limits. As the PD not only consists of a contribution from collimator-related radiation, but also of patient scatter, the differences are negligible when estimating the peripheral dose for an individual patient.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8635949     DOI: 10.1016/0360-3016(96)00083-1

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET.

Authors:  Vassiliki Vlachopoulou; Georgia Malatara; Harry Delis; Kiki Theodorou; Dimitrios Kardamakis; George Panayiotakis
Journal:  World J Radiol       Date:  2010-11-28

Review 2.  A review of dosimetry studies on external-beam radiation treatment with respect to second cancer induction.

Authors:  X George Xu; Bryan Bednarz; Harald Paganetti
Journal:  Phys Med Biol       Date:  2008-06-09       Impact factor: 3.609

3.  Immediate reconstruction with an expander/implant following ablatio mammae because of breast cancer : side effects and cosmetic results after adjuvant chest wall radiotherapy.

Authors:  Marc D Piroth; Daniela M Piroth; Michael Pinkawa; Seth G Woodruff; Richard Holy; Michael J Eble
Journal:  Strahlenther Onkol       Date:  2009-10-06       Impact factor: 3.621

4.  Consideration of the radiation dose delivered away from the treatment field to patients in radiotherapy.

Authors:  Michael L Taylor; Tomas Kron
Journal:  J Med Phys       Date:  2011-04

5.  Conversion coefficients for determination of dispersed photon dose during radiotherapy: NRUrad input code for MCNP.

Authors:  Mehrdad Shahmohammadi Beni; C Y P Ng; D Krstic; D Nikezic; K N Yu
Journal:  PLoS One       Date:  2017-03-31       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.