| Literature DB >> 8633766 |
T Solouki1, J A Marto, F M White, S Guan, A G Marshall.
Abstract
Significantly improved sensitivity for analysis of biomolecules by MALDI FT-ICR mass spectrometry is achieved by (i) microscope-monitored sample deposition onto a small indentation on the probe tip and (ii) multiple remeasurement of ions from a single laser shot. A simple modification to the solids probe tip allows for microdeposition of a few amols of analyte onto small indentation spots previously aligned with the laser beam. Ion multiple remeasurement of the same ion packet enhances the signal-to-noise ratio and thus extends the achievable FT-ICR MS detection limit. We demonstrate that FT-ICR can be used to detect parent and structurally significant fragment ions of peptides and phospholipids at low amol amounts. Positive ion mass spectra for approximately 90 amol of a mixture of angiotensin II and bradykinin, approximately 40 amol of dipalmitoylglycerophosphatidylcholine, and approximately 8 amol of substance P constitute the lowest reported detection limits to date for FT-ICR mass analysis of MALDI-generated ions.Entities:
Mesh:
Substances:
Year: 1995 PMID: 8633766 DOI: 10.1021/ac00118a017
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986