Mottle on computed radiographs of the chest in pediatric patients.

W Huda1, R M Slone, C J Belden, J L Williams, W A Cumming, C K Palmer.   

Abstract

PURPOSE: To investigate the relationship between radiation exposure and perceived mottle at bedside pediatric chest examinations performed with screen-film and computed radiographic techniques.
MATERIALS AND METHODS: In a pediatric intensive care unit, chest radiographs were obtained with both computed radiography (60 radiographs) and a 600-speed screen-film system (14 radiographs). The relative radiation exposure was estimated by using the sensitivity value obtained in the processing of each computed radiograph. Five radiologists assessed the mottle present in the computed radiographs and screen-film images.
RESULTS: For computed radiographs, the perceived level of mottle was inversely related to radiation exposure. For the same radiation exposure, the perceived mottle on computed radiographs was significantly higher than that on screen-film images (P < .001 for small cassettes; P < .01 for large cassettes).
CONCLUSION: Pediatric computed radiography of the chest requires approximately twice the exposure of a 600-speed screen-film system to attain the same level of mottle.

Entities:  

Mesh:

Year:  1996        PMID: 8633153     DOI: 10.1148/radiology.199.1.8633153

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  Optimization of dose and image quality for computed radiography and digital radiography.

Authors:  Charles E Willis
Journal:  J Digit Imaging       Date:  2007-03       Impact factor: 4.056

2.  Letter to the editor re: "Optimizing the use of computer radiography in pediatric chest imaging".

Authors:  Charles E Willis
Journal:  J Digit Imaging       Date:  2008-07-12       Impact factor: 4.056

3.  Challenges for pediatric radiology using computed radiography.

Authors:  C E Willis; B R Parker; M Orand; M L Wagner
Journal:  J Digit Imaging       Date:  1998-08       Impact factor: 4.056

4.  Support line and tube visibility in chest examinations using computed radiography.

Authors:  W Huda; C J Belden; L A Webb; C K Palmer
Journal:  J Digit Imaging       Date:  1997-08       Impact factor: 4.056

Review 5.  [Digital radiography of the chest in pediatric patients].

Authors:  S Puig
Journal:  Radiologe       Date:  2003-12       Impact factor: 0.635

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.