Patient-triggered ventilation decreases the work of breathing in neonates.

P H Jarreau1, G Moriette, P Mussat, C Mariette, A Mohanna, A Harf, H Lorino.   

Abstract

During conventional intermittent mandatory ventilation (IMV) in neonates, asynchrony between mechanical and spontaneous breaths is frequent. We tested the hypothesis that patient-triggered ventilation (PTV) reduces the work of breathing (WOB) by providing synchronized assistance for each breath. Accordingly, six intubated preterm infants were studied at the median postnatal age of 34 days while they were being weaned from mechanical ventilation (MV). Patients were ventilated using the Draeger Babylog 8000 (software #3) and studied in four successive modes of MV with a constant level of positive end-expiratory pressure. They were randomly assigned to IMV, PTV with peak inspiratory pressure of either 10 cm H2O (PTV10) or 15 cm H2O (PTV15), and spontaneous ventilation with continuous positive airway pressure. PTV was achieved in the assist/control mode. During PTV, infants adapted their pattern of breathing in response to an increase in tidal volume (median 7.5 ml/kg in IMV versus 8.2 in PTV10 and 8.5 in PTV15, p<0.05) by decreasing their respiratory rate, thus maintaining minute ventilation (439 ml/min/kg in IMV versus 422 in PTV10 and 455 in PTV15, NS) and transcutaneous CO2. WOB fell significantly during PTV compared with its level during IMV (0.81 J/L in IMV versus 0.48 and 0.47 during PTV10 and PTV15, respectively, p<0.05). Power of breathing decreased in the same proportions. These results demonstrate that PTV mode allows reduction of the workload imposed on the respiratory muscles.

Entities:  

Mesh:

Year:  1996        PMID: 8630564     DOI: 10.1164/ajrccm.153.3.8630564

Source DB:  PubMed          Journal:  Am J Respir Crit Care Med        ISSN: 1073-449X            Impact factor:   21.405


  10 in total

Review 1.  Weaning from assisted ventilation: art or science?

Authors:  S K Sinha; S M Donn
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2000-07       Impact factor: 5.747

2.  Work of breathing-tidal volume relationship: analysis on an in vitro model and clinical implications.

Authors:  G Natalini; D M Tuzzo; G Comunale; F A Rasulo; G Amicucci; A Candiani
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

3.  International randomised controlled trial of patient triggered ventilation in neonatal respiratory distress syndrome.

Authors:  J H Baumer
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2000-01       Impact factor: 5.747

Review 4.  Neonatal nasal intermittent positive pressure ventilation: what do we know in 2007?

Authors:  Louise S Owen; Colin J Morley; Peter G Davis
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2007-09       Impact factor: 5.747

Review 5.  New modes of mechanical ventilation in the preterm newborn: evidence of benefit.

Authors:  Nelson Claure; Eduardo Bancalari
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2007-09-05       Impact factor: 5.747

6.  Advances in respiratory support for high risk newborn infants.

Authors:  Eduardo Bancalari; Nelson Claure
Journal:  Matern Health Neonatol Perinatol       Date:  2015-05-21

7.  Simulation of pressure support for spontaneous breathing trials in neonates.

Authors:  Makoto Sasaki; Yoshikazu Yamaguchi; Tetsuya Miyashita; Yuko Matsuda; Masahide Ohtsuka; Osamu Yamaguchi; Takahisa Goto
Journal:  Intensive Care Med Exp       Date:  2019-02-08

Review 8.  Respiratory muscle function in the newborn: a narrative review.

Authors:  Theodore Dassios; Aggeliki Vervenioti; Gabriel Dimitriou
Journal:  Pediatr Res       Date:  2021-04-19       Impact factor: 3.953

9.  Design and Analysis of a Low-Cost Electronically Controlled Mobile Ventilator, Incorporating Mechanized AMBU Bag, for Patients during COVID-19 Pandemic.

Authors:  Rohan Lal Kshetry; Arnab Gupta; Somnath Chattopadhyaya; Madhulika Srivastava; Shubham Sharma; Jujhar Singh; Anirban Das Gupta; S Rajkumar
Journal:  J Healthc Eng       Date:  2022-03-30       Impact factor: 2.682

Review 10.  Synchronized mechanical ventilation for respiratory support in newborn infants.

Authors:  Anne Greenough; Thomas E Rossor; Adesh Sundaresan; Vadivelam Murthy; Anthony D Milner
Journal:  Cochrane Database Syst Rev       Date:  2016-09-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.