Tracheal gas insufflation-pressure control versus volume control ventilation. A lung model study.

H Imanaka1, R M Kacmarek, R Ritz, D Hess.   

Abstract

Tracheal gas insufflation (TGI) has been recommended as an adjunct to mechanical ventilation in the presence of elevated Pa CO2. Based on our initial clinical experience with continuous flow TGI and pressure control ventilation (PCV), we were concerned about elevation in peak airway pressure as TGI was applied. In a lung model, we evaluated the effects of continuous flow TGI during both PCV and volume control ventilation (VCV). A single compartment lung model was configured with an artificial trachea into which an 8-mm endotracheal tube was positioned. TGI was established with a 16-G catheter positioned 2 cm beyond the tip of the endotracheal tube. Ventilation was provided by a Puritan-Bennett 7200ae ventilator with PCV 20 cm H2O or VCV with a tidal volume (VTt) similar to that with PCV. A rate of 15 breaths/min and PEEP of 10 cm H2O were used throughout. Inspiratory times (TI) of 1.0, 1.5, 2.0, and 2.5 s were used with TGI of 0, 4, 8, and 12 L/min. Lung model compliance (ml/cm H2O) and resistance (cm H2O/L/s) combinations of 20/20, 20/5, and 50/20 were used. Auto-PEEP, VT, and peak alveolar and airway opening pressures increased as TGI and Ti increased, regardless of lung mechanics settings (p<0.01). All increases were greater with VCV than PCV (p<0.05). Continuous flow TGI with both PCV and VT-uncorrected VCV may result in marked increases in Vt and system pressures, especially at long TI.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8630540     DOI: 10.1164/ajrccm.153.3.8630540

Source DB:  PubMed          Journal:  Am J Respir Crit Care Med        ISSN: 1073-449X            Impact factor:   21.405


  3 in total

1.  A workable alternative to the problems with tracheal gas insufflation?

Authors:  R M Kacmarek
Journal:  Intensive Care Med       Date:  2002-08       Impact factor: 17.440

2.  Functional residual capacity measurement during tracheal gas insufflation.

Authors:  Y Fujino; M Nishimura; O Hirao; N Taenaka; I Yoshiya
Journal:  J Clin Monit Comput       Date:  1998-05       Impact factor: 2.502

3.  Expiratory automatic endotracheal tube compensation reduces dynamic hyperinflation in a physical lung model.

Authors:  Christoph Haberthür; Annekathrin Mehlig; John F Stover; Stefan Schumann; Knut Möller; Hans-Joachim Priebe; Josef Guttmann
Journal:  Crit Care       Date:  2009-01-23       Impact factor: 9.097

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.