Peripheral blood monocyte counting: towards a new reference method.

W Hübl1, S Andert, A Erath, A Lapin, P M Bayer.   

Abstract

Flow cytometric enumeration of monocytes stained with fluorescence-labelled monoclonal antibodies has been proposed as a possible reference method for monocyte counting. We compared precision and accuracy of monocyte counting of the Coulter STKS, the Cobas Argos 5 Diff, the 800-cell manual differential, and the Coulter Epics Profile II flow cytometer using double-staining with fluorescence-labelled monoclonal antibodies (CD45-FITC and CD14-PE). Precision: STKS, Argos and Profile II achieved a precision analogous to a 3423-, 1298-, and 11089-cell differential, respectively, confirming the superiority of automated methods. Accuracy (136 normal and abnormal samples): Correlation of automated methods with the manual differential was good (STKS: r = 0.934, Argos 5 Diff: r = 0.808, Profile II: r = 0.924; Spearman's rank correlation coefficient). The mean relative STKS monocyte result was 0.52 +/- 1.63% (mean +/- SD) higher than the manual differential, whereas the Argos 5 Diff results were 1.22 +/- 2.51% lower (p < 0.001). Profile II results showed a small bias against the manual differential (-0.18 +/- 1.44%, p < 0.05). Analysing 135 healthy adult subjects on the Profile II, males were found to have a higher mean monocyte count (relative count: 6.95 +/- 1.43% vs. 5.86 +/- 0.98%; absolute count: 0.48 +/- 0.15 x 10(9)/l vs. 0.39 +/- 0.11 x 10(9)/l, p < 0.001) and a higher and wider normal range than females (relative count: 4.97 to 9.78% vs. 4.26 to 7.81%, absolute count: 0.30 to 0.84 x 10(9)/l vs. 0.25 to 0.65 x 10(9)/l). Flow cytometry based on fluorescence-labelled monoclonal antibodies for monocyte enumeration seems an efficient tool to evaluate the monocyte counting performance of haematology analysers and an ideal successor to the manual differential as reference method for monocyte counting.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8620061     DOI: 10.1515/cclm.1995.33.11.839

Source DB:  PubMed          Journal:  Eur J Clin Chem Clin Biochem        ISSN: 0939-4974


  3 in total

1.  Properties of human blood monocytes. I. CD91 expression and log orthogonal light scatter provide a robust method to identify monocytes that is more accurate than CD14 expression.

Authors:  Dorothy Hudig; Kenneth W Hunter; W John Diamond; Doug Redelman
Journal:  Cytometry B Clin Cytom       Date:  2013-10-08       Impact factor: 3.058

2.  First-line chemoimmunotherapy in metastatic breast carcinoma: combination of paclitaxel and IMP321 (LAG-3Ig) enhances immune responses and antitumor activity.

Authors:  Chrystelle Brignone; Maya Gutierrez; Fawzia Mefti; Etienne Brain; Rosana Jarcau; Frédérique Cvitkovic; Nabil Bousetta; Jacques Medioni; Joseph Gligorov; Caroline Grygar; Manon Marcu; Frédéric Triebel
Journal:  J Transl Med       Date:  2010-07-23       Impact factor: 5.531

3.  Haemocompatibility of Modified Nanodiamonds.

Authors:  Michał Wąsowicz; Mateusz Ficek; Maciej S Wróbel; Ruchira Chakraborty; Dror Fixler; Paweł Wierzba; Małgorzata Jędrzejewska-Szczerska
Journal:  Materials (Basel)       Date:  2017-03-28       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.