Literature DB >> 861869

Thrombosis of Björk-Shiley aortic valve prosthesis: report of three cases.

A G de la Rocha, S K Plume, R J Baird.   

Abstract

Thrombotic malfunction of a Björk-Shiley aortic valve prosthesis occurred in three patients 6 to 16 months postoperatively. None of the patients had been taking anticoagulants. Although the presentation was acute, prodromal symptoms could be identified retrospectively in two of the patients. Two patients survived thrombectomy. Postoperative anticoagulant therapy is recommended in patients with these prostheses despite factors that may make such therapy riskier in specific patients. Attention to the character of murmurs and of the closure sound of the prosthetic valve must be part of the routine follow-up. In the emergency situation, when delay must be avoided, catheterization and angiography are unnecessary. The operative approach consists of complete thrombectomy without replacement of the valve or any of its components unless there is obvious periprosthetic leak or prosthetic wear.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 861869      PMCID: PMC1879496     

Source DB:  PubMed          Journal:  Can Med Assoc J        ISSN: 0008-4409            Impact factor:   8.262


  12 in total

1.  Long-term results of "simple" thrombectomy for thrombosed Björk-Shiley aortic valve prostheses.

Authors:  C L Byrd; W Z Yahr; J J Greenberg
Journal:  Ann Thorac Surg       Date:  1975-09       Impact factor: 4.330

2.  Encapsulation of the Björk-Shiley aortic disc valve prosthesis caused by the lack of anticoagulant treatment.

Authors:  V O Björk; A Henze
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1973

3.  Thrombosis on Bjork-Shiley aortic valve prosthesis. Clinical, arteriographic, echocardiographic and therapeutic observations in seven cases.

Authors:  J Ben-Zvi; F J Hildner; P A Chandraratna; P Samet
Journal:  Am J Cardiol       Date:  1974-10-03       Impact factor: 2.778

4.  Results of use of the pyrolytic carbon tilting disc Bjork-Shiley aortic prosthesis.

Authors:  J Fernandez; D Morse; V Maranhao; A S Gooch
Journal:  Chest       Date:  1974-06       Impact factor: 9.410

5.  Aortic valve replacement: new techniques, hydrodynamics, and clinical results.

Authors:  B J Messmer; G L Hallman; D Liotta; C Martin; D A Cooley
Journal:  Surgery       Date:  1970-12       Impact factor: 3.982

6.  Thrombosis encountered with Björk-Shiley prosthesis.

Authors:  A Y Bozer; A Karamehmetoglu
Journal:  J Cardiovasc Surg (Torino)       Date:  1972 Mar-Apr       Impact factor: 1.888

7.  The pyrolytic carbon occluder for the Björk-Shiley tilting disc valve prosthesis.

Authors:  V O Björk
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1972

8.  Clinical and haemodynamic results of aortic valve replacement with the Björk-Shiley tilting disc valve prosthesis.

Authors:  V O Björk; A Holmgren; C Olin; C O Ovenfors
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1971

9.  Surgical treatment of thrombosed Björk-Shiley aortic valve prosthesis.

Authors:  L A Gray; R L Fulton; T N Srivastava; N C Flowers
Journal:  J Thorac Cardiovasc Surg       Date:  1976-03       Impact factor: 5.209

10.  Sudden hemolysis indicating prosthetic valve dysfunction.

Authors:  J R Allard; G Fraser; A R Dobell
Journal:  Can J Surg       Date:  1976-05       Impact factor: 2.089

View more
  1 in total

1.  Thrombotic obstruction of the Björk-Shiley valve: the Glasgow experience.

Authors:  S J Ryder; H Bradley; J J Brannan; M A Turner; W H Bain
Journal:  Thorax       Date:  1984-07       Impact factor: 9.139

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.