Distortions in infant electrocardiograms caused by inadequate high-frequency response.

A S Berson, F Y Lau, J M Wojick, H V Pipberger.   

Abstract

Frank lead ECG's from infants were studied for frequency content by introducing low-pass filters of 50, 75, 100, and 150 150 Hz bandwidths before obtaining computer measurements. Results indicated that a minimum bandwidth of 100 Hz is required to avoid amplitude error of 10 per cent or greater. This bandwidth requierement is essentially the same as that required for adult ECG's despite the fact that infant QRS durations are usually about half those of adults. Although the average infant ECG spectrum is likely to contain higher frequencies than the average adult ECG spectrum, duration values for Q, R, and S waves overlap in these populations to such an extent that bandwidth requirements are practically identical.

Entities:  

Mesh:

Year:  1977        PMID: 860677     DOI: 10.1016/s0002-8703(77)80069-0