Comparison of three tonometers for measuring intraocular pressure in rabbits.

L S Abrams1, S Vitale, H D Jampel.   

Abstract

PURPOSE: Rabbits are used commonly for the evaluation of drugs and surgery to lower intraocular pressure (IOP). The authors compared the accuracy and variability of three tonometers for measuring IOP in rabbits.
METHODS: The anterior chamber of adult rabbits was cannulated with a 25-gauge needle connected to an elevated bottle of balanced salt solution. The bottle was raised and lowered to control IOP. A fluid-filled pressure transducer also was placed in the anterior chamber. Intraocular pressure was decreased in increments of 5 to 10 cm H2O from 50 to 0 mm Hg and was recorded with each of these instruments: the hand-held applanation tonometer, the Tono-pen XL, and the pneumatonometer.
RESULTS: The Tono-pen XL and the hand-held applanation tonometer underestimated the IOP, whereas the pneumatonometer slightly overestimated IOP. Under close stopcock conditions, at IOPs between 3 and 30 mm Hg, the Tono-pen XL was as accurate as the pneumatonometer but had a smaller variance; the hand-held applanation tonometer had lower accuracy and higher variability.
CONCLUSIONS: The Tono-pen XL is the tonometer of choice for measuring IOP in rabbits within the range of IOP 3 to 30 mm Hg. All tonometers were less accurate when the IOP was elevated markedly.

Entities:  

Mesh:

Year:  1996        PMID: 8603880

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  6 in total

1.  Diadenosine tetraphosphate induces tight junction disassembly thus increasing corneal epithelial permeability.

Authors:  P Loma; A Guzman-Aranguez; M J Pérez de Lara; J Pintor
Journal:  Br J Pharmacol       Date:  2014-12-15       Impact factor: 8.739

2.  Evaluation of rebound tonometry in non-human primates.

Authors:  Elizabeth J Elsmo; Julie A Kiland; Paul L Kaufman; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2011-02-16       Impact factor: 3.467

3.  Development of a Smart Pump for Monitoring and Controlling Intraocular Pressure.

Authors:  Simon A Bello; Sharad Malavade; Christopher L Passaglia
Journal:  Ann Biomed Eng       Date:  2016-09-27       Impact factor: 3.934

4.  Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen.

Authors:  Di Ma; Chong-Bo Chen; Jiajian Liang; Zhihao Lu; Haoyu Chen; Mingzhi Zhang
Journal:  Sci Rep       Date:  2016-10-12       Impact factor: 4.379

5.  Investigation of intraocular pressure of the anterior chamber and vitreous cavity of porcine eyes via a novel method.

Authors:  Koji Nagae; Hiromasa Sawamura; Makoto Aihara
Journal:  Sci Rep       Date:  2020-11-25       Impact factor: 4.379

6.  Strategies to estimate the characteristics of 24-hour IOP curves of treated glaucoma patients during office hours.

Authors:  Leonardo Colombo; Paolo Fogagnolo; Giovanni Montesano; Stefano De Cillà; Nicola Orzalesi; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2016-01-27       Impact factor: 2.209

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.