Literature DB >> 8601042

Carmeda surface heparinization in neonatal ECMO systems: long-term experiments in a sheep model.

K Palmer1, H Ehrén, R Benz, B Frenckner.   

Abstract

The thromboresistance in three Carmeda (Stockholm, Sweden) heparin-coated neonatal ECMO systems with a runtime of 45, 56 and 96 hours, respectively, and three noncoated systems with a runtime of 12, 42 and 66 hours, respectively, were compared using a sheep model. The flow rate was 200 ml/min and the activated clotting time (ACT) was kept at approximately 120 seconds. At the end of the experiment, the heparin-coated systems only contained minimal clotting while the controls showed major clotting in the entire system. Fibrin monomers were not detected until after 24 hours in the heparin-coated group, but demonstrated within 60 minutes in the noncoated group. It is concluded that the Carmeda heparin coating has a thromboresistant effect, and may be used to reduce the need for systemic heparinization in ECMO treatment of neonates.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8601042     DOI: 10.1177/026765919501000505

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  2 in total

Review 1.  Extracorporeal life support: experience with 2,000 patients.

Authors:  Brian W Gray; Jonathan W Haft; Jennifer C Hirsch; Gail M Annich; Ronald B Hirschl; Robert H Bartlett
Journal:  ASAIO J       Date:  2015 Jan-Feb       Impact factor: 2.872

Review 2.  Development and hemocompatibility testing of nitric oxide releasing polymers using a rabbit model of thrombogenicity.

Authors:  Terry C Major; Hitesh Handa; Gail M Annich; Robert H Bartlett
Journal:  J Biomater Appl       Date:  2014-06-16       Impact factor: 2.646