A comparative study of two near infrared spectrophotometers for the assessment of cerebral haemodynamics.

W N Colier1, N J van Haaren, B Oeseburg.   

Abstract

Conventional near infrared spectroscopy (NIRS), introduced by Jöbsis in 1977, can be considered as a reliable trend monitor for cerebral oxygenation. Quantisation, however, is complex and cumbersome. Recently a relatively simple system for cerebral oximetry (INVOS 3100, Somanetics Corporation, USA) was developed, measuring the regional oxygen saturation (rSO2) in the capillary bed of the cerebrum, presented as a numerical figure for easy interpretation. In this study a comparison was made between a conventional NIRS instrument and the new INVOS instrument, in order to obtain information about sensitivity and usefulness of the INVOS system. Changes in cerebral haemodynamics were induced by a moderate decrease of the arterial oxygen saturation (SaO2) and by varying the arterial carbon dioxide level (PaCO2). This will result in a higher (hypercapnia) or lower (hypocapnia) cerebral blood flow and subsequent change of both NIRS signals and INVOS signal. Healthy volunteers were used for this study. It was found that the steady state value for rSO2 was 70 +/- 6% (mean +/- SD). During the lowering of arterial saturation a poor correlation was found between rSO2 and SaO2 (r = 0.47). Increased cerebral blood flow induced by hypercapnia was detected by both conventional NIRS and the INVOS. Decreased cerebral blood flow induced by hypocapnia could only be detected by conventional NIRS. It was concluded that due to the variation in displayed rSO2 and the high amount of averaging in the algorithm the INVOS instrument does not yet provide more information than conventional NIRS.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8599259     DOI: 10.1111/j.1399-6576.1995.tb04342.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand Suppl        ISSN: 0515-2720


  5 in total

1.  The Effect of Parkinson Drug Timing on Cardiovascular Response during Treadmill Exercise in a Person with Parkinson Disease and Freezing of Gait.

Authors:  Cheryl M Petersen; Reid Nelson; Teresa M Steffen
Journal:  Physiother Can       Date:  2013       Impact factor: 1.037

2.  Optical fiber probe spectroscopy for laparoscopic monitoring of tissue oxygenation during esophagectomies.

Authors:  Daniel S Gareau; Frederic Truffer; Kyle A Perry; Kyle Perry; Thai H Pham; Thai Pham; C Kristian Enestvedt; James P Dolan; James Dolan; John G Hunter; Steven L Jacques
Journal:  J Biomed Opt       Date:  2010 Nov-Dec       Impact factor: 3.170

3.  Multichannel wearable fNIRS-EEG system for long-term clinical monitoring.

Authors:  Ali Kassab; Jérôme Le Lan; Julie Tremblay; Phetsamone Vannasing; Mahya Dehbozorgi; Philippe Pouliot; Anne Gallagher; Frédéric Lesage; Mohamad Sawan; Dang Khoa Nguyen
Journal:  Hum Brain Mapp       Date:  2017-10-23       Impact factor: 5.038

4.  Age-related deterioration of performance and increase of cortex activity comparing time- versus item-controlled fNIRS measurement.

Authors:  Leonore Blum; David Rosenbaum; Benjamin Röben; Katja Dehnen; Walter Maetzler; Ulrike Suenkel; Andreas J Fallgatter; Ann-Christine Ehlis; Florian G Metzger
Journal:  Sci Rep       Date:  2021-03-24       Impact factor: 4.379

5.  Parkinson's disease, lights and melanocytes: looking beyond the retina.

Authors:  Gregory L Willis; Cleo Moore; Stuart Maxwell Armstrong
Journal:  Sci Rep       Date:  2014-01-29       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.