Reliability of biologic indicators in a mail-return sterilization-monitoring service: a review of 3 years.

M T Andrés1, J M Tejerina, J F Fierro.   

Abstract

Most mail-return sterilization-monitoring services use spore strips to test sterilizers in dental clinics, but factors such as delay caused by mailing to the laboratory could cause false negatives. The aims of this study were to determine the influence of poststerilization time and temperature on the biologic indicator recovery system and to evaluate sterilization failure and its possible causes in dental clinics subscribing to a mail-return sterilization-monitoring service. Spore strips used in independent tests revealed the poststerilization time and temperature after a 7-day delay to have no significant influence. Sixty-six dental clinics that received quarterly biologic indicators to evaluate the effectiveness of their sterilizers had sterilization failure rates of 28.7% in 1992, 18.1% in 1993, and 9.1% in 1994, a statistically significant decrease in sterilization failure during the 3-year period. The usual causes of failure were operator error in wrapping of instruments, loading, operating temperature, or exposure time.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8596817

Source DB:  PubMed          Journal:  Quintessence Int        ISSN: 0033-6572            Impact factor:   1.677


  1 in total

1.  Effect of sodium hypochlorite and peracetic acid on the surface roughness of acrylic resin polymerized by heated water for short and long cycles.

Authors:  Felipe Sczepanski; Claudia Roberta Brunnquell Sczepanski; Sandrine Bittencourt Berger; Rafael Leonardo Xediek Consani; Alcides Gonini-Júnior; Ricardo Danil Guiraldo
Journal:  Eur J Dent       Date:  2014-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.