C L Bedwell1, D W Hamar, M L Hoesterey, J P Sonderman, K G Odde.
Abstract
Twenty forage samples were collected and selected for variation in nitrate content. Each forage samples was analyzed 4 times by 4 different methods: diphenylamine spot plate, spectrophotometric, nitrate-selective electrode, and high-performance liquid chromatographic. Five feed extracts were spiked with 2 different amounts of nitrate and analyzed by each method. The spectrophotometric and nitrate-selective electrode had similar percent recoveries, which were close to 100%. The nitrate-selective electrode method had the least variation of the 4 methods. The diphenylamine spot plate method had the poorest average recovery, greatest variation, and was the least accurate. The average coefficients of variation for all samples within a method were 15%, 12%, 6.4%, and 16 for the diphenylamine spot plate, spectrophotometric, nitrate-selective electrode, and high-performance liquid chromatographic methods, respectively. The variation in the nitrate-selective electrode method was lower (P < 0.05) than the other methods. The results from this study suggest that the nitrate-selective electrode method is more accurate and precise than the other methods of analysis tested.Entities:
Mesh:
Substances:
Year: 1995 PMID: 8580177 DOI: 10.1177/104063879500700418
Source DB: PubMed Journal: J Vet Diagn Invest ISSN: 1040-6387 Impact factor: 1.279