Nonlinear closed-loop control system for intracranial pressure regulation.

G L Coté1, R Durai, B Zoghi.   

Abstract

A nonlinear closed-loop control system with flat pressure-versus-flow characteristics that is aimed at regulating intracranial pressure (ICP) by adjusting the volume of cerebral spinal fluid (CSF) was designed, built, and tested. The control system design allows both the pressure setpoint and hysteresis to be adjusted to overcome the difficulties inherent in differential pressure-activated, fixed resistance, open-loop shunts. A dynamic six-compartment bench-top fluid system, which mimics the cerebral spinal fluid system, was designed, built, and tested. A computer simulation was developed which included the nonlinear on-off controller with hysteresis and a sixth-order, linear, multicompartmental model of the CSF system. The computer model and in vitro system results showed the ability of the system to track and compensate for pressure variations above and below normal as well as for spurious outputs that mimic such in vivo problems as blood pressure changes, sneezing, or coughing. There was one discrepancy between the simulated and in vitro results. The in vitro system had a higher rate of increase in pressure due to the more rigid compliance of the materials used, whereas the computer model compliance, based on the basal in vivo compliance of the CSF system, was less rigid. Based on these findings, the controller was modified to account for short-duration, extremely elevated pressures.

Mesh:

Year:  1995        PMID: 8572426     DOI: 10.1007/bf02584475

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  25 in total

1.  Compartmental analysis of compliance and outflow resistance of the cerebrospinal fluid system.

Authors:  A Marmarou; K Shulman; J LaMorgese
Journal:  J Neurosurg       Date:  1975-11       Impact factor: 5.115

2.  CSF hydrodynamic studies before and after cranioplasty.

Authors:  H Fodstad; J Ekstedt; H Fridén
Journal:  Acta Neurochir Suppl (Wien)       Date:  1979

3.  A critical analysis of valve shunts used in the treatment of hydrocephalus.

Authors:  S Hakim; F Duran de la Roche; J D Burton
Journal:  Dev Med Child Neurol       Date:  1973-04       Impact factor: 5.449

4.  Control of hydrocephalus by valve-regulated venous shunt: avoidance of complications in prolonged shunt maintenance.

Authors:  D P Becker; F E Nulsen
Journal:  J Neurosurg       Date:  1968-03       Impact factor: 5.115

5.  Intracranial pressure telemetry system using semicustom integrated circuits.

Authors:  A M Leung; W H Ko; T M Spear; J A Bettice
Journal:  IEEE Trans Biomed Eng       Date:  1986-04       Impact factor: 4.538

6.  Evaluation of a microsensor intracranial pressure transducer.

Authors:  S P Gopinath; L Cherian; C S Robertson; R K Narayan; R G Grossman
Journal:  J Neurosci Methods       Date:  1993-08       Impact factor: 2.390

7.  Closed-loop control of intracranial pressure.

Authors:  H L Rekate
Journal:  Ann Biomed Eng       Date:  1980       Impact factor: 3.934

8.  Complications of shunt operations performed in 206 children because of communicating hydrocephalus.

Authors:  Z A Huber
Journal:  Zentralbl Neurochir       Date:  1981

9.  Cerebrospinal fluid production by the choroid plexus and brain.

Authors:  T H Milhorat; M K Hammock; J D Fenstermacher; V A Levin
Journal:  Science       Date:  1971-07-23       Impact factor: 47.728

10.  Overdrainage phenomena in shunt treated hydrocephalus.

Authors:  K Faulhauer; P Schmitz
Journal:  Acta Neurochir (Wien)       Date:  1978       Impact factor: 2.216

View more
  1 in total

1.  New and improved ways to treat hydrocephalus: Pursuit of a smart shunt.

Authors:  Barry R Lutz; Pranav Venkataraman; Samuel R Browd
Journal:  Surg Neurol Int       Date:  2013-03-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.