Evaluation of a semi-automated platelet-counting system.

R M Rowan, C Fraser, J H Gray, G A McDonald.   

Abstract

Coulter Electronics Ltd have produced a semi-automated platelet-counting system. Platelet-rich plasma may be obtained either by tube sedimentation or by means of the Thrombo-fuge, the latter being an instrument designed to produce accelerated sedimentation. The instrument is linear over the entire range of platelet counts, and machine reproducibility is good. Comparison of machine-rated with visual counts satisfied statistical evaluation. The technique can be handled by one operator and platelet counts can be achieved at the rate of 30 per hour by both methods although individual counts on the Thrombo-fuge may be obtained in approximately one-quarter of the time required for tube sedimentation. The throughput using the Thrombo-fuge could certainly be doubled were two sample plates supplied. Few problems were encountered during the evaluation and most could be avoided by meticulous technique. Visual counts must be performed when the sample haematocrit is greater than 50%-Discrepant counts have been obtained in patients with white cell counts exceeding 50 X 10(9)/1 and in patients with giant platelets. ESR elevation for any reason does not lead to serious discrepancy in results. The incidence of platelet clumping due to the presence of platelet agglutinins and of microclot formation due to inadequate mixing is probably much higher than is commonly thought, and certainly peripheral blood film scrutiny should never be omitted in patients with low counts. Careful examination of peripheral blood films must be combined with instrument counting for some time lest further causes of discrepant counting emerge.

Entities:  

Mesh:

Year:  1977        PMID: 856881      PMCID: PMC476412          DOI: 10.1136/jcp.30.4.361

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  3 in total

1.  The reproducibility and constancy of the platelet count.

Authors:  G BRECHER; M SCHNEIDERMAN; E P CRONKITE
Journal:  Am J Clin Pathol       Date:  1953-01       Impact factor: 2.493

2.  Platelet counts with the Coulter counter.

Authors:  B S Bull; M A Schneiderman; G Brecher
Journal:  Am J Clin Pathol       Date:  1965-12       Impact factor: 2.493

3.  Evaluation of an automatic platelet counting system utilizing whole blood.

Authors:  R M Rowan; W Allan; R J Prescott
Journal:  J Clin Pathol       Date:  1972-03       Impact factor: 3.411

  3 in total
  2 in total

1.  A rapid and accurate differential centrifugation method for platelet counts.

Authors:  S Cousins; S M Lewis
Journal:  J Clin Pathol       Date:  1982-01       Impact factor: 3.411

2.  Use of theophylline in the investigation of pseudothrombocytopenia induced by edetic acid (EDTA-2K).

Authors:  O Ohnuma; Y Shirata; K Miyazawa
Journal:  J Clin Pathol       Date:  1988-08       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.