Literature DB >> 8567001

A comparison of adaptive and nonadaptive filters for reduction of power line interference in the ECG.

P S Hamilton1.   

Abstract

We have investigated the relative performance of an adaptive and nonadaptive 60-Hz notch filter applied to an ECG signal. We evaluated the performance of the two implementations with respect to adaptation rate (or transient response time), signal distortion, and implementation complexity. We also investigated the relative effect of adaptive and nonadaptive 60-Hz filtering on ECG data compression. With a 360 Hz sample rate and an adaptation time of approximately 0.3 s for a 1 mV 60-Hz signal, the adaptive implementation is less complex and introduces less noise, particularly in the ST-segment, into a typical ECG signal. When applied to ECG signals, prior to data compression by average beat subtraction and residual differencing, the residual signal resulting from the adaptively filtered signal had an average entropy 0.37 bits per sample (bps) lower than the unfiltered signal. The nonadaptive 60-Hz filter produced an average entropy decrease of 0.08 bps relative to the unfiltered ECG.

Mesh:

Year:  1996        PMID: 8567001     DOI: 10.1109/10.477707

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  13 in total

1.  A fast recursive-least-squares adaptive notch filter and its applications to biomedical signals.

Authors:  W K Ma; Y T Zhang; F S Yang
Journal:  Med Biol Eng Comput       Date:  1999-01       Impact factor: 2.602

2.  Trial-to-trial noise cancellation of cortical field potentials in awake macaques by autoregression model with exogenous input (ARX).

Authors:  Zheng Wang; Anna W Roe
Journal:  J Neurosci Methods       Date:  2010-11-11       Impact factor: 2.390

3.  Comparative study of ECG signal denoising by wavelet thresholding in empirical and variational mode decomposition domains.

Authors:  Salim Lahmiri
Journal:  Healthc Technol Lett       Date:  2014-09-16

4.  Genetic-algorithm cancellation of sinusoidal powerline interference in electrocardiograms.

Authors:  N Kumaravel; N Nithiyanandam
Journal:  Med Biol Eng Comput       Date:  1998-03       Impact factor: 2.602

5.  Deriving respiration from photoplethysmographic pulse width.

Authors:  Jesús Lázaro; Eduardo Gil; Raquel Bailón; Ana Mincholé; Pablo Laguna
Journal:  Med Biol Eng Comput       Date:  2012-09-21       Impact factor: 2.602

6.  Detection of multifiber neuronal firings: a mixture separation model applied to sympathetic recordings.

Authors:  Can Ozan Tan; J Andrew Taylor; Albert S H Ler; Michael A Cohen
Journal:  IEEE Trans Biomed Eng       Date:  2009-01       Impact factor: 4.538

7.  Removal of power-line interference from the ECG: a review of the subtraction procedure.

Authors:  Chavdar Levkov; Georgy Mihov; Ratcho Ivanov; Ivan Daskalov; Ivaylo Christov; Ivan Dotsinsky
Journal:  Biomed Eng Online       Date:  2005-08-23       Impact factor: 2.819

8.  Suppression of AC railway power-line interference in ECG signals recorded by public access defibrillators.

Authors:  Ivan Dotsinsky
Journal:  Biomed Eng Online       Date:  2005-11-26       Impact factor: 2.819

9.  Public access defibrillation: suppression of 16.7 Hz interference generated by the power supply of the railway systems.

Authors:  Ivaylo I Christov; Georgi L Iliev
Journal:  Biomed Eng Online       Date:  2005-03-15       Impact factor: 2.819

10.  Powerline noise elimination in biomedical signals via blind source separation and wavelet analysis.

Authors:  Samuel Akwei-Sekyere
Journal:  PeerJ       Date:  2015-07-02       Impact factor: 2.984

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.