Literature DB >> 8565238

Marked interference of hyperglycemia in measurements of mean (red) cell volume by Technicon H analyzers.

H L van Duijnhoven1, M Treskes.   

Abstract

Severe hyperglycemia can result in falsely high results for mean cell (erythrocyte) volume (MCV), which will also cause false results for erythrocyte indices calculated on the basis of MCV. Falsely high MCV results were obtained with the Technicon H1 and H2 analyzers and (to a lesser extent) with the Coulter T660. The H analyzers were more susceptible to this interference than was the Coulter T660. This difference in sensitivity of MCV to hyperglycemia can be explained by the use of sodium dodecyl sulfate in the Technicon erythrocyte diluent and by differences in incubation times. In severe hyperglycemia, results for MCV, mean cell hemoglobin concentration, and hematocrit obtained from electronic cell counters, especially Technicon H systems, are unreliable.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8565238

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  Elevated Hemoglobin and Macrocytosis: A Neglected Association to Become a Diagnostic Tool (A Case Report).

Authors:  Leonid L Yavorkovsky
Journal:  Perm J       Date:  2021-05-12
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.