Identification and quantification of hemoglobins A, F, S, and C by automated chromatography.

C Papadea1, J C Cate.   

Abstract

The Bio-Rad Variant Hemoglobin Testing System is an automated HPLC analyzer marketed with a Beta-thalassemia Short Program to quantify hemoglobins (Hbs) F and A2 and assist in detecting Hbs A, S, C, D, and E. We evaluated this system to replace several traditional methods for Hb in our hospital laboratory. Analytical performance relevant to quantifying Hbs A, S, C, and F was assessed with blood samples obtained from our local patient population. Studies of precision (CVs < 3%) and analytical limits (% of total Hb) of Hbs A (2-86%), F (1-89%), S (5-90%), and C (3-92%) demonstrated results comparable with or exceeding those of traditional methods. Results for patients' samples (n) for Hbs A (107), F (157), S (128), and C (27) correlated well (r > 0.93) with results by traditional methods. The satisfactory performance and efficiency led us to implement this system for routine quantification of clinically significant Hbs.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8565233

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

1.  Congenital erythropoietic porphyria due to a mutation in GATA1: the first trans-acting mutation causative for a human porphyria.

Authors:  John D Phillips; David P Steensma; Michael A Pulsipher; Gerald J Spangrude; James P Kushner
Journal:  Blood       Date:  2006-12-05       Impact factor: 22.113

2.  Thalassemia and iron deficiency in a group of northeast Thai school children: relationship to the occurrence of anemia.

Authors:  Nichathorn Panomai; Kanokwan Sanchaisuriya; Supawadee Yamsri; Pattara Sanchaisuriya; Goonnapa Fucharoen; Supan Fucharoen; Frank P Schelp
Journal:  Eur J Pediatr       Date:  2010-05-19       Impact factor: 3.183

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.