| Literature DB >> 8555565 |
H Harasaki1, K Fukamachi, A Massiello, J F Chen, S C Himley, F Fukumura, K Muramoto, S Niu, K Wika, C R Davies.
Abstract
A totally implantable, Cleveland Clinic-Nimbus total artificial heart (TAH) uses electrohydraulic energy conversion and an automatic left master-alternate mode control scheme, with a filling sensitivity of 1.0 l/min/mmHg and a maximum output of 9.5 l/min. The TAHs were tested in 12 calves for 1-120 days with normal major organ and blood cell function. Post-operative suppression of platelet aggregation recovered by the second post-operative week. The gelatin-coated pump surface generally was clean without any anticoagulants and free from infection. Embolism, which occurred in two cases, was caused by complications attributable to fungal infection in a Dacron graft and by thrombus formed around a jugular vein catheter. A system with a hybridized microcircuit controller in the interventricular space has been tested successfully in the three most recent cases, with a peak device surface temperature elevation of 6.5 degrees C. Heat effects were confined to the tissues immediately adjacent to the hottest spots. The carbon fiber-reinforced epoxy housing and 60 ml butyl rubber compliance chamber showed good tissue compatibility with a thin, fibrous tissue capsule. The transcutaneous energy transmission system and the internal battery functioned well as designed in the most recent animal implant.Entities:
Mesh:
Year: 1994 PMID: 8555565
Source DB: PubMed Journal: ASAIO J ISSN: 1058-2916 Impact factor: 2.872