On the potential of the Wilson central terminal with respect to an ideal reference for unipolar electrocardiography.

N Miyamoto1, Y Shimizu, G Nishiyama, S Mashima, Y Okamoto.   

Abstract

Body surface potential mapping was performed in 60 clinical cases and an ideal 0 potential was calculated in each case at 2msec intervals, which corresponds to the potential at infinity. Maximal deviation of the Wilson terminal voltage the ideal potential was 0.14mv on the average. Time course of potential variations of the central terminal was in proportion to the measured surface potentials. The lead vector of Wilson terminal was determined for each case, with a method to make a minimal difference between calculate and observed Wilson terminal voltage. The lead vector was directed superiorly and posteriorly with the magnitude of 24% of that of lead I on the average.

Mesh:

Year:  1995        PMID: 8551180     DOI: 10.1016/s0022-0736(05)80054-8

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  1 in total

1.  On the Einthoven Triangle: A Critical Analysis of the Single Rotating Dipole Hypothesis.

Authors:  Gaetano D Gargiulo; Paolo Bifulco; Mario Cesarelli; Alistair L McEwan; Hossein Moeinzadeh; Aiden O'Loughlin; Ibrahim M Shugman; Jonathan C Tapson; Aravinda Thiagalingam
Journal:  Sensors (Basel)       Date:  2018-07-20       Impact factor: 3.576