| Literature DB >> 8539166 |
M Yeşil1, S Bayata, N Postaci, C Aydin.
Abstract
A 17-year-old female with Kearns Sayre syndrome, complete heart block, and an implanted single chamber (VVIR) pacemaker, underwent testing with a GSM cellular phone that was placed directly over the pacemaker site. The pacemaker was immediately inhibited when the phone began to operate. A 6.5-second period of complete heart block with asystole occurred until the phone was switched off.Entities:
Mesh:
Year: 1995 PMID: 8539166 DOI: 10.1111/j.1540-8159.1995.tb03846.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976