The development and validation of a high performance liquid chromatography (HPLC)/tandem mass spectrometry assay for fenticonazole in human plasma and comparison with an HPLC-UV method.

W Speed1, J M Long, R J Simmonds, T A Enos.   

Abstract

A method is described for the determination of fenticonazole in human female plasma. The method utilizes high performance liquid chromatography coupled to atmospheric pressure positive-ion chemical ionization triple quadrupole mass spectrometry. Multiple reaction monitoring is employed for selectivity and sensitivity which enables quantification over the range 0.5-20 ng mL-1 with acceptable precision and accuracy. A comparison is made with an existing HPLC-UV assay and the utility of the technology of combined liquid chromatography and tandem mass spectometry for subnanogram per mL assays is discussed.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 8534893     DOI: 10.1002/rcm.1290091424

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  Determination of fenticonazole in human plasma by HPLC-MS/MS and its application to pharmacokinetic studies.

Authors:  Weixing Mao; Yiya Wang; Wenhui Hu; Feifei Jiao; Hongwei Fan; Li Ding
Journal:  J Pharm Anal       Date:  2016-09-08