Literature DB >> 8530778

Measuring the accuracy of the infrared tympanic thermometer: correlation does not signify agreement.

M Yaron1, S R Lowenstein, J Koziol-McLain.   

Abstract

This prospective study assessed the accuracy of the infrared tympanic thermometer (ITT) compared to the rectal thermometer (RT) using statistical measures of agreement. In a convenience sample of 100 adult emergency department patients, ear examinations to assess for cerumen or otitis were followed by temperature measurements using the First Temp 2000A thermometer in both ears and the IVAC 2000 rectally. Left and right ITT temperatures showed high correlation and agreement; therefore, only right ITT results are reported. Both the ITT and RT recorded similar mean temperatures, standard deviations, and ranges. The correlation of the ITT and RT and agreement were below the 0.8 level, indicating excellent agreement. The mean temperature difference (RT-ITT) between the two devices was 0.1 +/- 0.7 degrees C; in 10% of patients, the temperature difference was > or = 1 degree C. Among 10 patients identified as febrile by RT (RT > or = 38.5 degrees C), 6 were febrile by ITT. Significant differences occurred between the temperature measurements using the ITT and RT; these devices do not demonstrate excellent agreement.

Entities:  

Mesh:

Year:  1995        PMID: 8530778     DOI: 10.1016/0736-4679(95)00065-i

Source DB:  PubMed          Journal:  J Emerg Med        ISSN: 0736-4679            Impact factor:   1.484


  3 in total

Review 1.  The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis.

Authors:  Valentina Pecoraro; Davide Petri; Giorgio Costantino; Alessandro Squizzato; Lorenzo Moja; Gianni Virgili; Ersilia Lucenteforte
Journal:  Intern Emerg Med       Date:  2020-11-25       Impact factor: 3.397

2.  Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometer.

Authors:  Gasim I Gasim; Imad R Musa; Mohamed T Abdien; Ishag Adam
Journal:  BMC Res Notes       Date:  2013-05-10

3.  Studying the Accuracy and Function of Different Thermometry Techniques for Measuring Body Temperature.

Authors:  Aaron James Mah; Leili Ghazi Zadeh; Mahta Khoshnam Tehrani; Shahbaz Askari; Amir H Gandjbakhche; Babak Shadgan
Journal:  Biology (Basel)       Date:  2021-12-15