Roentgen stereophotogrammetric analysis for assessing migration of total hip replacement femoral components.

J Kiss1, D W Murray, A R Turner-Smith, C J Bulstrode.   

Abstract

A new Roentgen stereophotogrammetric analysis system, using a biplane technique, has been developed to determine the migration and rotation of total hip replacement (THR) femoral components in three dimensions. Stainless steel marker balls were injected into the femur during the operation. The patients stood within a calibration frame during the X-ray. The two exposures were taken consecutively allowing radio-opaque shutters to be moved in front of the films to prevent fogging. Studies with a model demonstrated that the system was capable of measuring the position of an implant to better than 0.11 mm (2 SD). In vivo measurements demonstrated that the migration rate of the different parts of the femoral component could be determined with an accuracy of 0.25 to 0.50 mm/year. By considering the accuracy determined in different ways, methods for improving the system have been identified. The migration and rotation rate of 58 Hinek cemented femoral components was studied for four years. Migration was three to five times greater (p < 0.001) during the first year than subsequently. The prosthesis head moved the most during the first year (0.94 mm). A better understanding of the cause of implant failure could be obtained by studying the early migration of different types of prosthesis and comparing this with their clinical results and design features.

Entities:  

Mesh:

Year:  1995        PMID: 8519406     DOI: 10.1243/PIME_PROC_1995_209_340_02

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  3 in total

1.  The John Charnley Award: Highly crosslinked polyethylene in total hip arthroplasty decreases long-term wear: a double-blind randomized trial.

Authors:  Siôn Glyn-Jones; Geraint E R Thomas; Patrick Garfjeld-Roberts; Roger Gundle; Adrian Taylor; Peter McLardy-Smith; David W Murray
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

2.  Prospective randomised clinical trial assessing subsidence and rotation, using radiostereometric analysis, of two modular cementless femoral stems (Global K2 and Apex).

Authors:  Mark Edmondson; Jay Ebert; Oscar Nivbrant; David Wood
Journal:  J Orthop       Date:  2014-04-22

3.  Notes on the use and interpretation of radiostereometric analysis.

Authors:  Brian Derbyshire; Robin J Prescott; Martyn L Porter
Journal:  Acta Orthop       Date:  2009-02       Impact factor: 3.717

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.