The performance of the Braunwald-Cutter aortic prosthetic valve.

E H Blackstone, J W Kirklin, J R Pluth, M E Turner, G V Parr.   

Abstract

Four hundred seventy-five patients underwent aortic valve replacement with the Braunwald-Cutter ball-valve prosthesis at two institutions. The early (30-day) hospital mortality was 4.7% for those with isolated aortic valve replacement and 6.9% for the entire group. For the former, 5-year actuarial survival of the hospital survivors was 72 +/- 5.7%; for the latter group it was 71 +/- 4.4%. Eleven patients (5 since the date of follow-up inquiry) have suffered poppet escape, 9 of whom died. The actuarial incidence of known poppet escape is 4 +/- 2.6% at 47 months; when the 5 patients suffering poppet escape since the date of follow-up inquiry are included, with certain assumptions, the incidence is 3.7 +/- 1.14%. The projected probability of poppet escape using all 11 patients is 12.2% at 5 years; the 70% confidence bands of projected probability of poppet escape separate from those of the risk of re-replacement at 61 months. This and other analyses indicate that in general, patients with the Braunwald-Cutter aortic prosthesis should have it replaced 4 1/2 to 5 years after its insertion.

Entities:  

Mesh:

Year:  1977        PMID: 849041     DOI: 10.1016/s0003-4975(10)64130-6

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  5 in total

1.  THE DAY THE LAST HEART VALVE WAS MADE.

Authors:  Duncan Barr
Journal:  Cardiovasc Dis       Date:  1978-06

2.  The Caged-Ball Prosthesis 60 Years Later: A Historical Review of a Cardiac Surgery Milestone.

Authors:  Andrea De Martino; Aldo D Milano; Mila Della Barbera; Gaetano Thiene; Uberto Bortolotti
Journal:  Tex Heart Inst J       Date:  2022-03-01

Review 3.  Bioengineering aspects of heart valve replacement.

Authors:  F J Schoen; J L Titus; G M Lawrie
Journal:  Ann Biomed Eng       Date:  1982       Impact factor: 3.934

4.  Normal function of a 43-year-old Braunwald Cutter heart valve.

Authors:  Gian Manuel Jiménez-Rodríguez; Sergio Criales-Vera; Mario Adrian Juárez-Peñaloza; Luis Alonso González-Tapia; Mariana Chaire-Hernández
Journal:  Oxf Med Case Reports       Date:  2018-02-26

5.  Mechanical prosthetic heart valves: here to stay!

Authors:  Sofia Kassou; Vassilios S Vassiliou
Journal:  Oxf Med Case Reports       Date:  2018-04-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.