Biomechanical analysis of a distally interlocked press-fit femoral total hip prosthesis.

N Mahomed1, J Schatzker, T Hearn.   

Abstract

This study was undertaken to compare the stability of a distally interlocked femoral prosthesis to a standard press-fit implant (Cementless Spotorno, Protek Co., Toronto, Canada). Five human femora were studied with each acting as its own control. All were tested in the locked mode followed by unlocked testing modes. The MTS 858 Biomaterials Testing System was used to test the implanted femoral stems under both pure torsional and pure axial loading. Distal interlocking increased the torsional stability by 320% (locked 4.05 Nm/degree; unlocked 1.25 Nm/degree) (P < .01) and the axial stability by 230% (locked 1,236.26 N/mm; unlocked 527.44 N/mm) (P < .01). the use of temporary distal interlock in cases where the initial stability of a press-fit implant is suboptimal should allow for improved biological fixation by minimizing micromotion at the stem-bone interface.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8478629     DOI: 10.1016/s0883-5403(06)80051-3

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Distal locking stem for revision femoral loosening and peri-prosthetic fractures.

Authors:  Patrice Mertl; Remy Philippot; Philippe Rosset; Henri Migaud; Jacques Tabutin; Denis Van de Velde
Journal:  Int Orthop       Date:  2010-12-24       Impact factor: 3.075

2.  Custom cementless stem improves hip function in young patients at 15-year followup.

Authors:  Xavier Flecher; Oliver Pearce; Sebastien Parratte; Jean-Manuel Aubaniac; Jean-Noel Argenson
Journal:  Clin Orthop Relat Res       Date:  2009-08-19       Impact factor: 4.176

3.  Results of uncemented distal locked prosthesis in revision hip arthroplasty with proximal femoral bone loss: A retrospective study.

Authors:  Sanjay Agarwala; Hari Ram Jhunjhunwala; Jawahir A Pachore; Siddharth B Joglekar; Krishna Kiran Eachempati
Journal:  Indian J Orthop       Date:  2013-01       Impact factor: 1.251

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.