G Flachenecker, K Fastenmeier, E Schmiedt, F Eisenberger.
Abstract
The cutting probe of transurethral resection devices has to be fed through a line inside of the shaft. The capacitive high-frequency currents between this feeder and the shaft flow through the urethra and may cause destructive heat within the urethra if a metallic shaft is used. This paper shows that no danger is caused from these capacitive currents as long as the capacity between feeder and shaft is less than 120 pF.Mesh: