Initial experience with a new flexible and adjustable mitral annuloplasty ring is described. Used in conjunction with standard techniques of mitral valve repair, this ring was implanted in 21 consecutive patients undergoing elective operation for mitral regurgitation. Satisfactory valve repair was feasible in 20 of these patients. Once the ring was sewn in place, adjustment of the ring to reduce or eliminate residual regurgitation was beneficial in 9 patients. Excellent results were achieved as determined by intraoperative transesophageal echocardiography. Comparable late results were found in all but 1 patient up to 1 year after repair. This patient required mitral valve replacement 6 months after repair due to failure of papillary muscle reconstruction. The BiFlex ring has the merits of a flexible prosthesis with the advantage of easy adjustment once secured in place.Entities:
Mesh:
Year: 1993 PMID: 8466339 DOI: 10.1016/0003-4975(93)90106-r